CB-X II DUAL-ENDED CLEANING BRUSH
Report
- Report Number
- 8030647-2009-00012
- Event Type
- Other
- Date Received
- September 18, 2009
- Date of Event
- August 19, 2009
- Report Date
- August 19, 2009
- Manufacturer
- AVENT S. DE R.L. DE C.V.
- Product Code
- FEB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ND, US
- Reporter Occupation
- OTHER
Narratives
THE INCIDENT SAMPLE HAS NOT BEEN PROVIDED FOR EVALUATION. REVIEW OF PULL TESTS RESULTS IN THE DEVICE HISTORY RECORD FOR THE LOT NUMBER PROVIDED FOUND AN AVERAGE OF 10.17LBS AND A STANDARD DEVIATION OF 0.45LBS, WITH THE LOWEST SPECIFICATION FOR THE ASSEMBLY AT 7LBS. IN ADDITION, SOME CHANGES TO THE MANUFACTURING PROCESS HAVE BEEN IMPLEMENTED SINCE (B)(6) 2009, WHICH INCLUDE THE USE OF A MAGNIFYING GLASS IN THE ASSEMBLY PROCESS AND REINFORCEMENTS OF THE GLUING PROCESS FOR THE TIP TO THE BRUSH. NO ADDITIONAL INFO HAS BEEN RECEIVED. WE WILL CONTINUE TO CLOSELY MONITOR THE FIELD PERFORMANCE OF THIS PRODUCT TO IDENTIFY EMERGING TRENDS AND IF APPLICABLE, ADDITIONAL INVESTIGATIONS AND CORRECTIVE ACTIONS WILL BE TAKEN.
KIMBERLY CLARK RECEIVED A COMPLAINT INDICATING THAT "A BROKEN TIP OF A CLEANING BRUSH FELL INTO THE PT". THE PIECE WAS TREATED AS A FOREIGN BODY AND REMOVED DURING THE PROCEDURE. THERE WAS NO HARM REPORTED TO THE PT. NO ADDITIONAL INFO HAS BEEN RECEIVED AT THIS TIME. KIMBERLY-CLARK HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFO THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE KIMBERLY-CLARK COMPLAINT DATABASE AND IDENTIFIED AS (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CB-X II DUAL-ENDED CLEANING BRUSH | ENDOSCOPIC CLEANING BRUSH | FEB | AVENT S. DE R.L. DE C.V. | M9043E302 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |