CB-X II DUAL-ENDED CLEANING BRUSH

MDR report key: 1480109 · Received September 18, 2009

Report

Report Number
8030647-2009-00012
Event Type
Other
Date Received
September 18, 2009
Date of Event
August 19, 2009
Report Date
August 19, 2009
Manufacturer
AVENT S. DE R.L. DE C.V.
Product Code
FEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INCIDENT SAMPLE HAS NOT BEEN PROVIDED FOR EVALUATION. REVIEW OF PULL TESTS RESULTS IN THE DEVICE HISTORY RECORD FOR THE LOT NUMBER PROVIDED FOUND AN AVERAGE OF 10.17LBS AND A STANDARD DEVIATION OF 0.45LBS, WITH THE LOWEST SPECIFICATION FOR THE ASSEMBLY AT 7LBS. IN ADDITION, SOME CHANGES TO THE MANUFACTURING PROCESS HAVE BEEN IMPLEMENTED SINCE (B)(6) 2009, WHICH INCLUDE THE USE OF A MAGNIFYING GLASS IN THE ASSEMBLY PROCESS AND REINFORCEMENTS OF THE GLUING PROCESS FOR THE TIP TO THE BRUSH. NO ADDITIONAL INFO HAS BEEN RECEIVED. WE WILL CONTINUE TO CLOSELY MONITOR THE FIELD PERFORMANCE OF THIS PRODUCT TO IDENTIFY EMERGING TRENDS AND IF APPLICABLE, ADDITIONAL INVESTIGATIONS AND CORRECTIVE ACTIONS WILL BE TAKEN.

Description of Event or Problem · 1

KIMBERLY CLARK RECEIVED A COMPLAINT INDICATING THAT "A BROKEN TIP OF A CLEANING BRUSH FELL INTO THE PT". THE PIECE WAS TREATED AS A FOREIGN BODY AND REMOVED DURING THE PROCEDURE. THERE WAS NO HARM REPORTED TO THE PT. NO ADDITIONAL INFO HAS BEEN RECEIVED AT THIS TIME. KIMBERLY-CLARK HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFO THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE KIMBERLY-CLARK COMPLAINT DATABASE AND IDENTIFIED AS (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CB-X II DUAL-ENDED CLEANING BRUSH ENDOSCOPIC CLEANING BRUSH FEB AVENT S. DE R.L. DE C.V. M9043E302

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention