PSI 140*100*40 PEEK
Report
- Report Number
- 8030965-2022-04256
- Event Type
- Malfunction
- Date Received
- June 23, 2022
- Date of Event
- May 26, 2022
- Report Date
- June 23, 2022
- Manufacturer
- SYNTHES GMBH
- Product Code
- GXN
- UDI-DI
- 07611819341357
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
PRODUCT COMPLAINT # (B)(4). DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. COMPLAINANT PART IS NOT EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE NOT STERILE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THE PHOTO WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE PHOTO REVEALED THAT THE PSI 140*100*40 PEEK HAD NO SIGNS OF ISSUE OR DEFECT. DIMENSIONAL DISCREPANCY CANNOT BE CONFIRMED SINCE THE DEVICE WAS NOT RETURNED, THEREFORE, A DIMENSIONAL INSPECTION COULD NOT BE PERFORMED. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. THE OVERALL COMPLAINT WAS NOT CONFIRMED FOR THE PSI 140*100*40 PEEK. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION AND HENCE THE ROOT CAUSE CANNOT BE DETERMINED. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES.
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SINGAPORE AS FOLLOWS: IT WAS REPORTED THAT DURING A CRANIOPLASTY SURGERY, THE PEEK IMPLANT HAS A GAP AT THE FRONTAL REGIONAL. THIS COMPLAINT INVOLVES ONE (1) DEVICE PSI 140*100*40 PEEK. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 761939 | PSI 140*100*40 PEEK | PLATE, CRANIOPLASTY, PREFORMED, NON-ALTERABLE | GXN | SYNTHES GMBH | 819P791 | 07611819341357 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |