FDA Adverse Event Malfunction Summary report: N

THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S

MDR report key: 14776793 · Received June 22, 2022

Report

Report Number
8010047-2022-10552
Event Type
Malfunction
Date Received
June 22, 2022
Date of Event
May 26, 2022
Report Date
May 3, 2024
Manufacturer
AOMORI OLYMPUS CO., LTD.
Product Code
GEI
UDI-DI
04953170383571
PMA / PMN Number
K211838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

CUSTOMER REPORTED FOUR EVENTS OF FAILED THUNDERBEAT DEVICES THAT OCCURRED IN FROM MARCH TO MAY 2022, THE ISSUE BEING PROBE BROKEN OFF. THESE EVENTS ARE CAPTURED IN MEDWATCHES WITH PATIENT IDENTIFIERS (B)(6) (TB-0520FCS, 1JB154), (B)(6) (TB-0535FCS, PW307904), (B)(6) (TB-0535FCS, KR179419), AND (B)(6) (TB-0535FCS, PW307861). THIS MEDWATCH IS FOR THE PATIENT IDENTIFIER C22214516. THE DEVICE IS NOT RETURNED, AS SUCH A DEFINITIVE ROOT CAUSE OF THE REPORTED COMPLAINT CANNOT BE DETERMINED AT THIS TIME. THIS EVENT IS UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR UPON RECEIVING ADDITIONAL INFORMATION.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE DEVICE EVALUATION AND THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. PLEASE SEE THE UPDATES IN SECTIONS D9, H3, H4, H6, AND H10. THE DEVICE WAS RETURNED AND AN EVALUATION WAS COMPLETED FOR IT. DURING INSPECTION, OLYMPUS CONFIRMED THE REPORTED EVENT, THE PROBE WAS BROKEN OFF. IT WAS DISCOVERED THAT CRACKS WERE PROGRESSING FROM THE SCRATCHED OF THE PROBE. ADDITIONALLY, THERE WAS WEAR OF THE TISSUE PAD AND PEELING OF THE COATING ON THE GRIP; HOWEVER, THESE DEFECTS ARE NOT CONSIDERED SEVERE ENOUGH TO CAUSE A POTENTIAL ADVERSE EVENT. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. ALTHOUGH A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED, THE FOLLOWING STEP-BY-STEP SCENARIO LIKELY CAUSED THE EVENT: 1. AS THE VICINITY OF THE ORIGIN OF THE FRACTURE SURFACE ARE BLACKENED, A CRACK WAS GENERATED IN THE PROBE DUE TO A LOAD SUCH AS A SPARK . 2. THE CRACK THEN EXTENDED WHEN LOAD TO THE DEVICE WAS APPLIED TO THE PROBE. AND THE PROBE BROKE OFF IN THE END. THE FOLLOWING INFORMATION IS INCLUDED IN THE INSTRUCTIONS FOR USE (IFU): ¿SHOULD ANY IRREGULARITY (ERROR WINDOW, ABNORMAL NOISE, ABNORMAL OUTPUT, ABNORMAL OPERATION, ABNORMAL APPEARANCE, ETC.) OR MALFUNCTION BE OBSERVED WHILE USING THE THUNDERBEAT, STOP THE USE AND WITHDRAW THE INSTRUMENTS FROM THE BODY CAVITY. DO NOT WITHDRAW THE TRANSDUCER PLUG FROM THE ULTRASONIC GENERATOR. CHECK THE EQUIPMENT BY REFERRING TO CHAPTER 8, ¿TROUBLESHOOTING¿, IN THE INSTRUCTION MANUAL FOR THE ULTRASONIC GENERATOR. IF THE IRREGULARITY REMAINS AFTER TROUBLESHOOTING, REPLACE THE TRANSDUCER. IF THIS DOES NOT RESOLVE THE ISSUE, REPLACE THE THUNDERBEAT INSTRUMENT. IF THE IRREGULARITY REMAINS UNRESOLVED, CONTACT OLYMPUS.¿ ¿USE THE THUNDERBEAT INSTRUMENT PROPERLY. IMPROPER USE MAY CAUSE THE PROBE TIP TO FALL OFF INSIDE THE BODY CAVITY, PREMATURE WEAR, PARTIAL SEPARATING, DEFORMATION, BREAKAGE, PATIENT AND/OR OPERATOR INJURY, MALFUNCTION OF THE CARDIAC PACEMAKER, BURNS OF THE SURGEON, SURGICAL STAFF AND/OR PATIENT, ELECTRIC SHOCK, ABNORMAL OUTPUT OR MALFUNCTION, PERFORATION, BLEEDING, POSTOPERATIVE BLEEDING, TISSUE DAMAGE AND INFECTION TO THE SURGEON, SURGICAL STAFF AND/OR PATIENT.¿ ¿IF THE GRASPING SECTION, METAL-EXPOSED AREA AROUND IT OR THE PROBE TIP GETS STICKED TISSUE DURING TREATMENT, WIPE IT WITH A SOFT OBJECT SUCH AS A PIECE OF GAUZE OR A BRUSH. DO NOT ATTEMPT TO SCRAPE IT WITH A SHARP OBJECT SUCH AS A SCALPEL OR THE TIP OF TWEEZERS. OTHERWISE, THE GRASPING SECTION, METAL-EXPOSED AREA AROUND IT, THE FLUORINE RESIN PART, A COATED SURFACE OR THE PROBE TIP MAY BE SCRATCHED AND DAMAGED, WHICH MAY LEAD TO FALL-OFF OF THE DAMAGED PART INTO THE BODY CAVITY OR BURNS OF THE TISSUE BY A HIGH-FREQUENCY LEAK CURRENT OUTPUT DUE TO DESTRUCTION OF THE INSULATION STRUCTURE.¿ OLYMPUS WILL CONTINUE TO MONITOR THE PERFORMANCE OF THIS DEVICE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO CORRECT THE LEGAL MANUFACTURER¿S CONTACT INFORMATION AND FACILITY REGISTRATION NUMBER. THE FACILITY REGISTRATION NUMBER IS (B)(4).

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. PLEASE SEE THE UPDATES IN SECTIONS H6 AND H10. THE DEVICE WAS NOT RETURNED; THEREFORE, THE REPORTED PHENOMENON OR CONDITION OF THE DEVICE COULD NOT BE CONFIRMED. HOWEVER, IT IS POSSIBLE TO INFER THE CAUSE FROM THE RESULTS OF PAST SIMILAR INVESTIGATIONS. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT THE PROBE BROKE DUE TO CONTACT WITH A SURGICAL INSTRUMENT OR THE NON-INSULATED AREA OF GRASPING SECTION. ALTHOUGH A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED, THE FOLLOWING STEP-BY-STEP SCENARIO LIKELY CAUSED THE EVENT: 1. DURING OUTPUT IN SEAL & CUT MODE, THE PROBE CAME IN CONTACT WITH HARD TISSUE, METAL OR A SURGICAL INSTRUMENT. CONSEQUENTLY, A SCRATCH WAS MADE ON THE PROBE. 2. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 3. THE PROBE BROKE WHEN ADDED LOAD. OR 1. THE DISTAL END OF THE TISSUE PAD WAS WORN AWAY BECAUSE ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED WHILE GRASPING NOTHING (INCLUDING THE CASE THE USER KEPT ACTIVATING AFTER CUTTING TISSUE). 2. THE TISSUE PAD WAS WORN AWAY, CAUSING THE NON-INSULATED OF THE GRASPING SECTION TO TOUCH THE PROBE. 3. ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED UNDER THIS SITUATION, THEN THE SCRATCHES INDICATING THAT THE PROBE AND GRASPING SECTION WERE IN CONTACT WITH EACH OTHER WERE MADE. 4. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 5. THE PROBE BROKE WHEN ADDED LOAD. THIS FOLLOWING INFORMATION IS INCLUDED IN THE INSTRUCTIONS FOR USE (IFU): ¿DO NOT ACTIVATE OUTPUT IN SEAL & CUT MODE WHILE THE GRASPING SECTION IS CLOSED WITHOUT CONTACTING TISSUE OR VESSEL, OR ENSURING THAT TISSUE IS TRANSECTED. OTHERWISE, A LOCAL INCREASE OF THE TEMPERATURE DUE TO A FRICTION BETWEEN THE PROBE TIP AND THE GRASPING SECTION MAY RESULT IN VARIOUS FORMS OF DAMAGE IN THE PROBE TIP AND/OR THE TISSUE PAD, SUCH AS PREMATURE WEAR, BREAKAGE, DEFORMATION, AND/OR FALLING OFF INSIDE THE BODY CAVITY AND/OR PARTIAL SEPARATING.¿ ¿WHEN CUTTING AND VESSEL SEALING IS PERFORMED IN SEAL & CUT MODE, APPLY LIGHT TENSION ON THE TISSUE SO THAT USERS CAN CONFIRM IT IS TRANSECTED. ALSO, STOP ACTIVATION IMMEDIATELY AFTER TISSUE IS TRANSECTED. OTHERWISE, THE GRASPING SECTION, THE TISSUE PAD, OR THE PROBE TIP MAY BREAK AND FALL OFF, AND PARTIAL SEPARATING OF THE TISSUE PAD MAY OCCUR DUE TO A LOCAL INCREASE OF TEMPERATURE CAUSED BY THE FRICTION BETWEEN TISSUE PAD AND THE PROBE TIP DURING ACTIVATION.¿ ¿THE THUNDERBEAT INSTRUMENT SHOULD BE USED FOR SOFT TISSUE. DO NOT ACTIVATE OUTPUT WHILE GRASPING HARD TISSUE SUCH AS BONE OR HIGHLY CALCIFIED TISSUE, OR HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR, FORCEPS, AND OTHERS). OTHERWISE, IT MAY CAUSE THE PROBE TIP TO BE SCRATCHED OR COME INTO DIRECT CONTACT WITH THE METAL AREA OF THE GRASPING SECTION AS THE HEAT GENERATED BY THE FRICTION BETWEEN THE HARD OBJECT AND THE PROBE TIP COULD CAUSE WEAR/DEFORMING/SPLITTING/PROTRUDING/PARTIAL SEPARATING OF THE TISSUE PAD. IN TURN, THE PROBE MAY BREAK BEFORE DISPLAYING AN ERROR WINDOW OR GENERATING AN ALARM TONE.¿ ¿DO NOT GRASP OR LET THE PROBE TIP CONTACT HARD OBJECTS SUCH AS METAL CLIPS, STAPLER, OR OTHER INSTRUMENTS (E.G., UTERINE MANIPULATOR). ALSO, BE CAREFUL TO AVOID CONTACTING THE PROBE TIP WITH THOSE ACCIDENTALLY. PARTICULARLY DURING ACTIVATION, A SCRATCH ON THE PROBE TIP COULD OCCUR DUE TO ULTRASONIC VIBRATION, WHICH LEADS THE PROBE TIP TO BREAK AND FALL OFF INTO THE BODY CAVITY. IN ADDITION, THE HIGH-FREQUENCY (RF BIPOLAR) CURRENT FLOWS THROUGH THE METAL AND GENERATES SPARK DISCHARGE, WHICH MAY CAUSE BURNS AND DECREASE FUNCTIONALITIES.¿ OLYMPUS WILL CONTINUE TO MONITOR THE PERFORMANCE OF THIS DEVICE.

Description of Event or Problem · 0

AS REPORTED FOR THIS EVENT BY THE CUSTOMER, DURING A LIVER PROCEDURE THE DEVICE PROBE BROKE OFF AND WAS RETRIEVED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A NEW SIMILAR DEVICE. PER THE NURSE, THE DEVICE DID NOT COME IN CONTACT WITH OTHER INSTRUMENTS OR SPREADERS, AND THERE WAS NO MISHANDLING OF THE DEVICE BY THE SURGEON THAT CONTRIBUTED TO THE EVENT. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308119 THUNDERBEAT 5 MM, 20 CM, FRONT-ACTUATED GRIP TYPE S ULTRASONIC SURGICAL DEVICE GEI AOMORI OLYMPUS CO., LTD. TB-0520FCS 1JB154 04953170383571

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown