MONOMAX VIOLET 2/0 (3) 90CM HR30 (M)
Report
- Report Number
- 3003639970-2022-00292
- Event Type
- Malfunction
- Date Received
- June 22, 2022
- Report Date
- July 27, 2022
- Manufacturer
- B. BRAUN SURGICAL, S.A.
- Product Code
- NWJ
- PMA / PMN Number
- K100876
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
IF ADDITIONAL INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SUBMITTED.
ANALYSIS AND RESULTS: THERE ARE NO PREVIOUS COMPLAINTS OF THIS CODE-BATCH. WE MANUFACTURED AND DISTRIBUTED IN THE MARKET (B)(4) OF THIS CODE-BATCH. THERE ARE NO UNITS IN OUR STOCK. WE HAVE RECEIVED 21 CLOSED SAMPLES FOR ANALYSIS. TIGHTNESS TEST TO THE SAMPLES RECEIVED HAS BEEN PERFORMED AND THE UNITS ARE TIGHT. WE HAVE TESTED THE NEEDLE ATTACHMENT STRENGTH OF ALL SAMPLES RECEIVED AND THE RESULTS FULFIL THE REQUIREMENTS OF THE EUROPEAN PHARMACOPOEIA (EP): 1.63 KGF IN AVERAGE AND 1.37 KGF IN MINIMUM (EP REQUIREMENTS: 1.12 KGF IN AVERAGE AND 0.46 KGF IN MINIMUM) REVIEWED THE BATCH MANUFACTURING RECORD, THERE ARE NO INCIDENCES REPORTED REGARDING THIS ISSUE AND THE RESULTS DURING THE PROCESS FULFIL USP/EP AND B. BRAUN SURGICAL REQUIREMENTS. FINAL CONCLUSION: ALTHOUGH THE RESULTS OF THE SAMPLES RECEIVED FULFIL THE SPECIFICATIONS OF EUROPEAN PHARMACOPOEIA/ B. BRAUN SURGICAL SPECIFICATIONS, WE TAKE NOTE OF THIS INCIDENCE IN ORDER TO ASSESS IF NEW OR ADDITIONAL ACTIONS ARE NEEDED. WE REGRET ANY INCONVENIENCE THIS ISSUE MAY HAVE CAUSED AND THANK YOU FOR YOUR COLLABORATION. ACTIONS ON PRODUCT DISTRIBUTED OF THIS REFERENCE/BATCH: BASED ON THE CONCLUSION DERIVED FROM INVESTIGATION, IT IS NOT REQUIRED TO MAKE ACTIONS IN DISTRIBUTED PRODUCT. CORRECTIVE/PREVENTIVE ACTIONS: ACCORDING TO OUR INTERNAL PROCEDURES, THERE IS NO NEED TO ESTABLISH CORRECTIVE OR PREVENTIVE ACTIONS. NEVERTHELESS, THIS COMPLAINT IS RECORDED FOR TRENDING ANALYSIS TO ASSESS IF ACTIONS ARE NEEDED IN THE FUTURE.
IT WAS REPORTED AN ISSUE WITH MONOMAX SUTURE. THE CLIENT REPORTED THAT THE NEEDLE COMES OFF THE THREAD. NO PATIENT INVOLVEMENT, OCCURRED PRIOR TO USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 599191 | MONOMAX VIOLET 2/0 (3) 90CM HR30 (M) | SYNTHETIC ABSORBABLE MONOFILAR SUTURE | NWJ | B. BRAUN SURGICAL, S.A. | B0041278 | 121461 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |