MAGNIFYING LPE 2.5X RED R 12IN TO 15IN WRKG DIST
Report
- Report Number
- 3014334038-2022-00124
- Event Type
- Malfunction
- Date Received
- June 21, 2022
- Report Date
- June 21, 2022
- Manufacturer
- INTEGRA LIFESCIENCES MANSFIELD
- Product Code
- HJH
- UDI-DI
- 10381780462453
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE MAGNIFYING LOUPE, 2.5X RED FRAME, 12IN TO 15IN WORKING DISTANCE (1-5001) WAS RETURNED FOR EVALUATION. DEVICE HISTORY RECORD: NO ABNORMALITIES RELATED TO THE REPORTED ISSUE WERE IDENTIFIED. FAILURE ANALYSIS: EVALUATION OF THE RETURNED MAGNIFYING LOUPE FOUND THE DEVICE WAS IN USED CONDITION WITH THE ADJUSTMENT KNOBS DISCONNECTED. ROOT CAUSE ANALYSIS: THE COMPLAINT REPORTED BY THE CUSTOMER WAS CONFIRMED. THE RETURNED MAGNIFYING LOUPE HAD ITS ADJUSTMENT KNOB BROKEN OFF DUE TO ROUGH HANDLING/ENVIRONMENTAL DAMAGE. NO MANUFACTURING, WORKMANSHIP OR MATERIAL DEFICIENCY HAS BEEN IDENTIFIED. NO FURTHER INVESTIGATION REQUIRED BASED ON THE ACCEPTABILITY OF RISK AND NO ADVERSE TRENDS IDENTIFIED. THIS WILL BE MONITORED AND TRENDED GOING FORWARD.
IT WAS REPORTED THAT THE UP AND DOWN MECHANISM OF THE MAGNIFYING LOUPE (1-5001) WAS NOT WORKING. NO PATIENT INJURY, DEATH OR SURGICAL DELAY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2043797 | MAGNIFYING LPE 2.5X RED R 12IN TO 15IN WRKG DIST | M5 - GENERAL SURGERY | HJH | INTEGRA LIFESCIENCES MANSFIELD | 1-5001 | W2102005 | 10381780462453 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |