FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ 3-PIECE SYRINGE

MDR report key: 14748381 · Received June 20, 2022

Report

Report Number
3003152976-2022-00282
Event Type
Malfunction
Date Received
June 20, 2022
Date of Event
May 24, 2022
Report Date
July 27, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR LOT 2203076, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. RETAINED SAMPLES OF THE SAME LOT WERE USED FOR ADDITIONAL EVALUATION. THE PRODUCT WAS VISUALLY INSPECTED AND NO DEFECTS WERE IDENTIFIED. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE AND LOT SIZE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES. GIVEN NO ISSUES WERE OBSERVED WITH THE RETAINED SAMPLES AND THE DEVICE HISTORY RECORDS DO NOT INDICATE ANY ISSUES RELATED TO THIS INCIDENT, WE CANNOT IDENTIFY A ROOT CAUSE AT THIS TIME.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 3-PIECE SYRINGE SCALE MARKING ARE LIGHT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE ARE ONLY A FEW GRADUATION MARKS WITH LIGHT (ALMOST NOT VISIBLE) PRINTING.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE BD PLASTIPAK¿ 3-PIECE SYRINGE SCALE MARKING ARE LIGHT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THERE ARE ONLY A FEW GRADUATION MARKS WITH LIGHT (ALMOST NOT VISIBLE) PRINTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1090149 BD PLASTIPAK¿ 3-PIECE SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2203076

Patients

Seq Age Sex Outcome Treatment
1 Unknown