FDA Adverse Event Malfunction Summary report: N

ID NOW COVID-19 ASSAY

MDR report key: 14748299 · Received June 20, 2022

Report

Report Number
1221359-2022-02956
Event Type
Malfunction
Date Received
June 20, 2022
Date of Event
June 1, 2022
Report Date
July 6, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QJR
UDI-DI
10811877011269
PMA / PMN Number
EUA2000074
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION IS STILL IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 1060011 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR KIT PART NUMBER 190-000 / LOT 1060011 AND TEST BASE PART NUMBER 190-430 / LOT 1060011. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE POSITIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 1060011 SHOWED THAT THE COMPLAINT RATE IS (B)(4)%. ABBOTT DIAGNOSTICS SCARBOROUGH WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE. HOWEVER A POSSIBLE ASSIGNABLE ROOT CAUSE IS SAMPLE INTERFERENCE OR CROSS CONTAMINATION.

Description of Event or Problem · 0

THE CUSTOMER REPORTED A FALSE POSITIVE REPORT WITH THE ID NOW COVID-19 ASSAY ON (B)(6) 2022. THE PATIENT WAS INITIALLY TESTED WITH THE ID NOW COVID-19 ASSAY WHICH GENERATED A POSITIVE RESULT. CONFIRMATION PCR TESTING (PLATFORM UNKNOWN) WAS PERFORMED ON A NASAL SWAB SAMPLE ON (B)(6) 2022. THE PATIENT WAS REPORTED TO BE MILDLY SYMPTOMATIC AND WAS QUARANTINING. NO FURTHER INFORMATION REGARDING PATIENT HEALTH STATUS OR OUTCOME WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2254124 ID NOW COVID-19 ASSAY MOLECULAR IVD FOR ID NOW COVID-19 QJR ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 1060011 10811877011269

Patients

Seq Age Sex Outcome Treatment
1 Unknown