BD¿ LUER SLIP SYRINGE STERILE, SINGLE USE, 10 ML
Report
- Report Number
- 1213809-2022-00346
- Event Type
- Malfunction
- Date Received
- June 20, 2022
- Date of Event
- May 23, 2022
- Report Date
- July 8, 2022
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
H6: INVESTIGATION SUMMARY: THREE PHOTOS OF 10ML SLIP-TIP SYRINGES (P/N 303134) BATCH 1013522 IN A BLISTERPACK WERE RECEIVED AND EVALUATED. ONE IMAGE SHOWS TWO SEALED PACKAGES WITH THE TOP WEB GRAPHICS SHOWING ALL APPLICABLE INFORMATION. OTHER TWO IMAGES SHOW A BLACK BEETLE BUG INSECT THE SIZE OF THE PLUNGER ROD THUMBREST OBSERVED OUTSIDE OF THE FLUID PATH LAYING ON TOP OF THE PLUNGER ROD THUMBREST. ONE IMAGE SHOWS A CLOSE-UP OF THE BEETLE BUG AND THE OTHER IMAGE SHOWS TWO SEALED PACKAGES NEXT TO ONE ANOTHER WITH THE INSECT IN ONE OF THEM WITH MANY BLACK PARTICULATES, COMPOSITION OF FOREIGN MATTER CANNOT BE DETERMINED BASED UPON THE PHOTOS RECEIVED. THE BLISTER PACKAGE WAS SEALED IN THE PHOTOS AND PHYSICALLY DID NOT HAVE ANY TEARS OR HOLES INDICATING THE INSECT HAD WORKED ITS WAY INTO THE PACKAGE. THE SYRINGE RECEIVED WAS NON-CONFORMING PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE FOREIGN MATTER DEFECT MAY BE ASSOCIATED WITH THE MATERIAL HANDLING PROCESS. SINCE THE INSECT WAS INSIDE THE SYRINGE OUTSIDE OF THE FLUID PATH IT IS MOST LIKELY ENTERED AFTER ASSEMBLY BUT PRIOR TO THE IN-LINE PACKAGING PROCESS. BD CANAAN MAINTAINS A ROBUST PEST CONTROL PROGRAM THAT INCLUDES STRATEGICALLY PLACED INSECT LIGHT TRAPS THROUGHOUT THE FACILITY. INTERIOR INSECT CONTROL IN COMBINATION WITH EXTERIOR INSECT CONTROL BY A THIRD PARTY HELPS TO ENSURE THE MAXIMUM AMOUNT OF CONTAINMENT FOR ANY PESTS THAT MAY ATTEMPT TO ENTER THE FACILITY. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 1013522. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
IT WAS REPORTED THAT THE BD¿ LUER SLIP SYRINGE STERILE, SINGLE USE, 10 ML EXPERIENCED FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE FOUND A BEETLE PACKAGED INSIDE THE BD 10ML SLIP TIP SYRINGES.
IT WAS REPORTED THAT THE BD¿ LUER SLIP SYRINGE STERILE, SINGLE USE, 10 ML EXPERIENCED FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: WE HAVE FOUND A BEETLE PACKAGED INSIDE THE BD 10ML SLIP TIP SYRINGES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2814295 | BD¿ LUER SLIP SYRINGE STERILE, SINGLE USE, 10 ML | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 1013522 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |