TOWEL OR BLUE STERILE
Report
- Report Number
- 1423537-2022-00792
- Event Type
- Malfunction
- Date Received
- June 17, 2022
- Date of Event
- May 18, 2022
- Report Date
- June 17, 2022
- Manufacturer
- CARDINAL HEALTH 200, LLC
- Product Code
- FRL
- UDI-DI
- 10630140917257
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- OTHER
Narratives
BASED ON SUPPLIER INVESTIGATION, DEVICE HISTORY RECORD (DHR) REVIEW DID NOT INDICATE ANY EXCEPTION THAT COULD LEAD TO THE REPORTED INCIDENT. THE AVERAGE LINTING DATA IS 0.105G/10PCS. A SAMPLE WAS NOT RETURNED FOR INVESTIGATION. ACCORDING TO SUPPLIER, THE OR TOWEL IS MADE OF COTTON, SO COTTON FIBER IS BORN. SUPPLIER IS CONTINUOUSLY WORKING WITH CARDINAL HEALTH TO BETTER CONTROL THE LINTING AND HAVE IMPLEMENTED SEVERAL MEASURES TO IMPROVE IT: SUCTION MACHINES HAVE BEEN INSTALLED IN GREY CLOTH ROLLING PROCESS, DYEING PROCESS AND CUTTING PROCESS, THE SUCTION PROCESS WAS ADDED BEFORE PRODUCT'S FINAL FOLDING, AND WORKERS DO IT ACCORDING TO STANDARD OPERATION PROCEDURE REQUIREMENT, LINTING TEST METHOD AND ACCEPTABLE CRITERIA WAS STIPULATED TO SEE THE SUCTION RESULTS. (=0.38G/10PIECES), AND IN THE FOLDING PROCESS, SUPPLIER USED ONE CLOTH PAD UNDER 100 PIECES SEMI-FINISHED PRODUCTS TO AVOID LINTING STUCK ONTO THE PRODUCTS DURING PRODUCT'S TRANSFER. FROM THE INVESTIGATION, THERE IS NO ABNORMAL SITUATION HAPPENED IN PRODUCTION OR DHR. THEREFORE, THE ROOT CAUSE COULD NOT BE DETERMINED. CARDINAL WILL CONTINUE TO MONITOR TRENDS FOR THIS PRODUCT.
CUSTOMER REPORTED THE OR TOWELS WERE DISINTEGRATING AFTER GETTING WET IN THE STERILE SALINE IN THE STERILE BOWLS, DURING A WATCHMAN LEFT ATRIAL APPENDAGE OCCLUSION DEVICE PROCEDURE. FUZZY TYPE PIECES WERE COMING OFF. THERE WAS NO HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 203580 | TOWEL OR BLUE STERILE | FIBER, MEDICAL, ABSORBENT | FRL | CARDINAL HEALTH 200, LLC | 28700-006 | 220106-18-SH | 10630140917257 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Male |