Report

Report Number
1423537-2022-00792
Event Type
Malfunction
Date Received
June 17, 2022
Date of Event
May 18, 2022
Report Date
June 17, 2022
Manufacturer
CARDINAL HEALTH 200, LLC
Product Code
FRL
UDI-DI
10630140917257
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BASED ON SUPPLIER INVESTIGATION, DEVICE HISTORY RECORD (DHR) REVIEW DID NOT INDICATE ANY EXCEPTION THAT COULD LEAD TO THE REPORTED INCIDENT. THE AVERAGE LINTING DATA IS 0.105G/10PCS. A SAMPLE WAS NOT RETURNED FOR INVESTIGATION. ACCORDING TO SUPPLIER, THE OR TOWEL IS MADE OF COTTON, SO COTTON FIBER IS BORN. SUPPLIER IS CONTINUOUSLY WORKING WITH CARDINAL HEALTH TO BETTER CONTROL THE LINTING AND HAVE IMPLEMENTED SEVERAL MEASURES TO IMPROVE IT: SUCTION MACHINES HAVE BEEN INSTALLED IN GREY CLOTH ROLLING PROCESS, DYEING PROCESS AND CUTTING PROCESS, THE SUCTION PROCESS WAS ADDED BEFORE PRODUCT'S FINAL FOLDING, AND WORKERS DO IT ACCORDING TO STANDARD OPERATION PROCEDURE REQUIREMENT, LINTING TEST METHOD AND ACCEPTABLE CRITERIA WAS STIPULATED TO SEE THE SUCTION RESULTS. (=0.38G/10PIECES), AND IN THE FOLDING PROCESS, SUPPLIER USED ONE CLOTH PAD UNDER 100 PIECES SEMI-FINISHED PRODUCTS TO AVOID LINTING STUCK ONTO THE PRODUCTS DURING PRODUCT'S TRANSFER. FROM THE INVESTIGATION, THERE IS NO ABNORMAL SITUATION HAPPENED IN PRODUCTION OR DHR. THEREFORE, THE ROOT CAUSE COULD NOT BE DETERMINED. CARDINAL WILL CONTINUE TO MONITOR TRENDS FOR THIS PRODUCT.

Description of Event or Problem · 0

CUSTOMER REPORTED THE OR TOWELS WERE DISINTEGRATING AFTER GETTING WET IN THE STERILE SALINE IN THE STERILE BOWLS, DURING A WATCHMAN LEFT ATRIAL APPENDAGE OCCLUSION DEVICE PROCEDURE. FUZZY TYPE PIECES WERE COMING OFF. THERE WAS NO HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203580 TOWEL OR BLUE STERILE FIBER, MEDICAL, ABSORBENT FRL CARDINAL HEALTH 200, LLC 28700-006 220106-18-SH 10630140917257

Patients

Seq Age Sex Outcome Treatment
1 Male