FDA Adverse Event Malfunction Summary report: N

BD 5ML SYRINGE SLIP TIP WITH BD PRECISIONGLIDE¿ NEEDLE 22G 1 1/4

MDR report key: 14734256 · Received June 17, 2022

Report

Report Number
3002682307-2022-00169
Event Type
Malfunction
Date Received
June 17, 2022
Date of Event
May 19, 2022
Report Date
July 19, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMI
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H6: INVESTIGATION SUMMARY A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED MATERIAL NUMBER (B)(4) AND LOT NUMBER 1904152. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. AS SAMPLES WERE UNAVAILABLE FOR RETURN, TWENTY RETAINED SAMPLES WERE OBTAINED. THROUGH EVALUATION OF THE RETAINED SAMPLES, NO SIGNS OF DEFECT COULD BE OBSERVED. BASED ON THE LIMITED INVESTIGATION RESULTS AND THE PROVIDED CUSTOMER FEEDBACK, WE ARE UNABLE TO CONFIRM NOR DETERMINE AN EXACT CAUSE FOR THIS REPORTED ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD 5ML SYRINGE SLIP TIP WITH BD PRECISIONGLIDE¿ NEEDLE 22G 1 1/4 FOREIGN MATTER WAS FOUND IN THE FLUID PATHWAY. THIS OCCURRED 5 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: OPERATING ROOM THE NURSE IN CHARGE FOUND A WHITE FOREIGN MATTER ON THE INNER WALL OF THE SYRINGE DURING SAMPLING.

Description of Event or Problem · 0

IT WAS REPORTED WHILE USING BD 5ML SYRINGE SLIP TIP WITH BD PRECISIONGLIDE¿ NEEDLE 22G 1 1/4 FOREIGN MATTER WAS FOUND IN THE FLUID PATHWAY. THIS OCCURRED 5 TIMES. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: OPERATING ROOM THE NURSE IN CHARGE FOUND A WHITE FOREIGN MATTER ON THE INNER WALL OF THE SYRINGE DURING SAMPLING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1311132 BD 5ML SYRINGE SLIP TIP WITH BD PRECISIONGLIDE¿ NEEDLE 22G 1 1/4 HYPODERMIC SINGLE LUMEN NEEDLE FMI BECTON DICKINSON, S.A. 1904152