BD PLASTIPAK¿ SYRINGES
Report
- Report Number
- 3003152976-2022-00274
- Event Type
- Malfunction
- Date Received
- June 17, 2022
- Date of Event
- April 17, 2022
- Report Date
- June 1, 2022
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- FMF
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT 2201089, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. FINAL PRODUCTS IN THIS MANUFACTURING LINE, FOR THIS REFERENCE ARE SAMPLED AND THEY ARE SUBJECTED TO VISUAL AND FUNCTIONAL INSPECTIONS DURING THE DIFFERENT MANUFACTURING SUB-PROCESSES ACCORDING TO PROCEDURES, NO ISSUES RELATED TO THE REPORTED INCIDENT WERE IDENTIFIED DURING MANUFACTURING. DAMAGE CAN OCCUR DUE TO THE PRODUCT JAMMING WITHIN MANUFACTURING EQUIPMENT. WITHOUT THE SAMPLE TO EVALUATE, WE CANNOT VERIFY THIS TO BE TRUE FOR THE INCIDENT REPORTED. BASED ON THE AVAILABLE INFORMATION WE CANNOT IDENTIFY A ROOT CAUSE AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL BE MONITORED BY OUR QUALITY TEAM FOR SIGNS OF EMERGING TRENDS.
IT WAS REPORTED WHILE USING BD PLASTIPAK¿ SYRINGES LEAKAGE OCCURRED FROM A HOLE. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DATE OF OCCURRENCE: (B)(6) 2022. INCIDENT DESCRIPTION: A DEPARTMENT REPORTED TO US A LEAK OF NOREPINEPHRINE SOLUTION DURING SYRINGE PREPARATION. ON VISUAL INSPECTION CLEARLY SHOWS THAT THE BODY OF THE SYRINGE WAS ALREADY DAMAGED AND THAT THIS SYRINGE SHOULD HAVE BEEN IMMEDIATELY DISCARDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1351358 | BD PLASTIPAK¿ SYRINGES | PISTON SYRINGE | FMF | BECTON DICKINSON, S.A. | 2201089 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |