FDA Adverse Event Injury Summary report: N

STEM: AMISTEM H 01.18.133 HA COATED STD STEM SIZE 3

MDR report key: 14733655 · Received June 17, 2022

Report

Report Number
3005180920-2022-00457
Event Type
Injury
Date Received
June 17, 2022
Date of Event
May 3, 2022
Report Date
September 22, 2022
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LZO
UDI-DI
07630030804090
PMA / PMN Number
K093944
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEW PERFORMED ON 19 MAY 2022. LOT 110842: 60 ITEMS MANUFACTURED AND RELEASED ON 13-APR-2011. EXPIRATION DATE: 2016-MAR-31. NO ANOMALIES FOUND RELATED TO THE PROBLEM. SINCE RELEASE TO DATE, 57 ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT.

Additional Manufacturer Narrative · 0

VISUAL INSPECTION PERFORMED BY R&D MANAGER: DURING THE ANALYSIS IT IS EVALUATED THAT APPROXIMATELY THE FIRST HALF OF THE STEM BODY IS STILL COVERED WITH HA COATING. THIS IN UNEXPECTED, MEANING THAT THE STEM WAS NOT CORRECTLY ONSTEOINTEGRATED WITH PATIENT BONE. SOME SINGS AND SCRATCHES ARE PRESENT ON THE STEM NECK PROBABLY DUE TO REVISION SURGERY. IT IS NOT POSSIBLE TO DETERMINE THE ROOT CAUSE OF REPORTED STEM LOOSENING.

Description of Event or Problem · 0

PRIMARY SURGERY PERFORMED ON 23 MAY 2011. THE PATIENT WAS REPORTING PAIN SINCE OCTOBER 2021. ABOUT 11 YEARS AFTER THE PRIMARY SURGERY, STEM WAS REVISED DUE TO ASEPTIC LOOSENING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1350225 STEM: AMISTEM H 01.18.133 HA COATED STD STEM SIZE 3 HIP COATED STEM LZO MEDACTA INTERNATIONAL SA 01.18.133 110842 07630030804090

Patients

Seq Age Sex Outcome Treatment
1 67 YR Male Required Intervention