FDA Adverse Event Death Summary report: N

670G INSULIN PUMP MMT-1780KL

MDR report key: 14728661 · Received June 17, 2022

Report

Report Number
2032227-2022-214772
Event Type
Death
Date Received
June 17, 2022
Date of Event
December 28, 2021
Report Date
November 30, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000166526
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(4). S.W VERSION 4.10E, RETAINER RING = CLEAR. THE CUSTOMER PASSED AWAY ON (B)(6) 2021. ALSO, ON (B)(4), SVN #: (B)(4) - CUSTOMER RETURNED PUMP FOR AN ALLEGED TEST PLUG DAMAGE FOUND ON (B)(6) 2019. UNABLE TO CONFIRM TEST PLUG DAMAGE DUE TO PUMP RECEIVED WITHOUT THE ORIGINAL TEST PLUG. THE PUMP PASSED THE FUNCTIONAL TESTS, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND DAT AT 0.08715 INCHES. THE PUMP WAS PROGRAMMED WITH A TEST GUARDIAN LINK (3) TRANSMITTER AND A GLUCOSE SENSOR SIMULATOR. THE PUMP CONNECTED SUCCESSFULLY TO THE TRANSMITTER AND DISPLAYED ¿TRANSMITTER CONNECTION SUCCESSFUL¿. NO UNEXPECTED SENSOR ERRORS OR CONNECTION ANOMALIES NOTED. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS. SUCCESSFULLY UPLOADED PUMP TO CARELINK. THE TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT. HOWEVER, A CRACKED RETAINER WAS NOTED DURING TESTING. THE FOLLOWING WERE ALSO NOTED DURING VISUAL INSPECTION: A CRACKED KEYPAD OVERLAY, A PILLOWING KEYPAD OVERLAY AND A SCRATCHED CASE. THE PUMP WAS RECEIVED WITHOUT A BATTERY INSTALLED. PLEASE SEE BELOW FOR THE DATE RANGE LISTED IN THE FORMATTED HISTORY FILE. THE FORMATTED HISTORY FILE LISTS DATA FROM (B)(6) 2021 & (B)(6) 2022. A CRACKED RETAINER WAS CONFIRMED. UNABLE TO CONFIRM TEST PLUG DAMAGE DUE TO PUMP RECEIVED WITHOUT THE ORIGINAL TEST PLUG. THE PUMP PASSED ALL THE REQUIRED TESTING. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Additional Manufacturer Narrative · 0

THIS MDR RELATED TO THE PUERTO RICO MANUFACTURING SITE HAS BEEN ASSIGNED A MEDWATCH NUMBER FROM THE MEDTRONIC MINIMED NORTHRIDGE SITE, PER VARIANCE 5. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. THE INFORMATION THAT PROVIDED WITH THE INITIAL REPORT WAS INCORRECT. THE CORRECT INFORMATION HAS BEEN INCLUDED WITH THIS REPORT IN B5, D10 AND H6 SECTION. THE HEALTH EFFECT - IMPACT CODE, MEDICAL DEVICE PROBLEM CODE AND COMPONENT CODE THAT WAS PROVIDED WITH THE INITIAL REPORT WAS INCOMPLETE. THE COMPLETE INFORMATION HAS BEEN INCLUDED WITH THIS REPORT IN H6 SECTION.

Description of Event or Problem · 0

IT WAS REPORTED VIA PHONE CALL THAT THE CUSTOMER PASSED AWAY. THE CAUSE OF DEATH WAS CARDIAC ARREST. THE CUSTOMER DID NOT WORN THE INSULIN PUMP AT THE TIME OF DEATH. CUSTOMER WAS NOT USING THE SENSOR. CUSTOMER PASSED AWAY ON (B)(6) 2021. THE CALLER WILL RETURN THE INSULIN PUMP FOR ANALYSIS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162451 670G INSULIN PUMP MMT-1780KL ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KL HG3J3BV 000000763000166526

Patients

Seq Age Sex Outcome Treatment
1 69 YR Male Death FRN-MMT-332-RSVR, UNOMED-MMT-387A