FDA Adverse Event Malfunction Summary report: N

X3C 2200 WITH MAGELLAN 3.4 SOFTWARE INSATLLED

MDR report key: 1470095 · Received July 30, 2009

Report

Report Number
9616853-2009-00001
Event Type
Malfunction
Date Received
July 30, 2009
Date of Event
July 2, 2009
Report Date
July 27, 2009
Manufacturer
IMAGING DYNAMICS COMPANY (IDC)
Product Code
KPR
PMA / PMN Number
K071401
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Description of Event or Problem · 1

AN IMAGING DYNAMICS COMPANY (IDC) X3C 2200 DIGITAL RADIOGRAPHIC SYSTEM WAS IN USE AT (B)(6). AT APPROX 9:00 PM AN IMAGE OF THE PT TAKEN BY THE RADIOLOGY TECHNICIAN APPEARED ON THE SYSTEM PREVIEW SCREEN, BUT SUBSEQUENTLY DID NOT APPEAR IN THE STUDY ON THE IDC UNIT. THE SUBJECT REQUIRED AN ADDITIONAL EXPOSURE TO COMPENSATE FOR THE MISSING IMAGE. UPON FURTHER INVESTIGATION, IT WAS DETERMINED THE COMPUTER WAS NOT CONFIGURED AS REQUIRED BY ICD, AND A LACK OF ADEQUATE AVAILABLE COMPUTER MEMORY CAUSED A LONG IMAGE PROCESSING TIME TO OCCUR. DURING THE INITIAL IMAGE PROCESSING, THE RADIOLOGY TECHNICIAN ATTEMPTED TO TAKE ANOTHER EXPOSURE CAUSING THE INITIAL IMAGE TO BE LOST. THERE WAS NO INJURY TO ANY HEALTH PROFESSIONAL. THERE WAS NO ACCIDENTAL RADIATION EXPOSURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 X3C 2200 WITH MAGELLAN 3.4 SOFTWARE INSATLLED DIGITAL X-RAY IMAGING SYSTEM KPR IMAGING DYNAMICS COMPANY (IDC) X3C 2200 (M3. 4)

Patients

Seq Age Sex Outcome Treatment
1