X3C 2200 WITH MAGELLAN 3.4 SOFTWARE INSATLLED
Report
- Report Number
- 9616853-2009-00001
- Event Type
- Malfunction
- Date Received
- July 30, 2009
- Date of Event
- July 2, 2009
- Report Date
- July 27, 2009
- Manufacturer
- IMAGING DYNAMICS COMPANY (IDC)
- Product Code
- KPR
- PMA / PMN Number
- K071401
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
AN IMAGING DYNAMICS COMPANY (IDC) X3C 2200 DIGITAL RADIOGRAPHIC SYSTEM WAS IN USE AT (B)(6). AT APPROX 9:00 PM AN IMAGE OF THE PT TAKEN BY THE RADIOLOGY TECHNICIAN APPEARED ON THE SYSTEM PREVIEW SCREEN, BUT SUBSEQUENTLY DID NOT APPEAR IN THE STUDY ON THE IDC UNIT. THE SUBJECT REQUIRED AN ADDITIONAL EXPOSURE TO COMPENSATE FOR THE MISSING IMAGE. UPON FURTHER INVESTIGATION, IT WAS DETERMINED THE COMPUTER WAS NOT CONFIGURED AS REQUIRED BY ICD, AND A LACK OF ADEQUATE AVAILABLE COMPUTER MEMORY CAUSED A LONG IMAGE PROCESSING TIME TO OCCUR. DURING THE INITIAL IMAGE PROCESSING, THE RADIOLOGY TECHNICIAN ATTEMPTED TO TAKE ANOTHER EXPOSURE CAUSING THE INITIAL IMAGE TO BE LOST. THERE WAS NO INJURY TO ANY HEALTH PROFESSIONAL. THERE WAS NO ACCIDENTAL RADIATION EXPOSURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | X3C 2200 WITH MAGELLAN 3.4 SOFTWARE INSATLLED | DIGITAL X-RAY IMAGING SYSTEM | KPR | IMAGING DYNAMICS COMPANY (IDC) | X3C 2200 (M3. 4) |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |