FDA Adverse Event Injury Summary report: N

PRIMARY MANDIBULAR RECONSTRUCTION PLATE, 11-HOLES, WITH TEMPLATE

MDR report key: 14697610 · Received June 15, 2022

Report

Report Number
0008010177-2022-00018
Event Type
Injury
Date Received
June 15, 2022
Date of Event
May 20, 2022
Report Date
August 30, 2022
Manufacturer
STRYKER LEIBINGER FREIBURG
Product Code
JEY
UDI-DI
04546540356376
PMA / PMN Number
K014263
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE HAS NOT YET BEEN RECEIVED AT THE MANUFACTURER FOR TESTING. AN EVALUATION WILL BE CONDUCTED UPON RECEIPT OF THE DEVICE, AND A FOLLOW-UP REPORT WILL BE SUBMITTED AFTER THE QUALITY INVESTIGATION IS COMPLETE. IN ADDITION TO THE PRIMARY MANDIBULAR RECONSTRUCTION PLATE, 11-HOLES, WITH TEMPLATE (55-15911) THERE WERE ALSO A TOTAL QUANTITY OF 5 REPORTED SCREWS (50-20412 AND 50-20408 ) WHICH ARE CAPTURED IN COMPLAINTS # (B)(4).

Additional Manufacturer Narrative · 0

THE REPORTED EVENT ¿BREAKAGE - POSTOPERATIVE¿ COULD BE CONFIRMED, BECAUSE OF THE VISUAL INSPECTION. THE MACROSCOPIC APPEARANCE POINTS TO EXCEEDING THE MATERIAL STRENGTH. THAT IS CONFIRMED BY THE DETERIORATION OF THE ANODIZATION LAYER, THE CHANGES OF THE SURFACE¿S STRUCTURE AND THE SECONDARY CRACKS IN THE AREA OF THE BREAKAGE. STRONG PLASTIC DEFORMATIONS WITH MATERIAL BULGING AND A METAL COIL INDICATE HIGH FORCES ACTED ON THE PLATE DURING APPLICATION / REMOVAL. THE LOT NUMBERS ARE NOT AVAILABLE, AND NO POSTOPERATIVE CT-SCANS OR X-RAYS OF THE INITIAL SURGERY CAN BE PROVIDED. THE SENIOR GLOBAL MEDICAL DIRECTOR STATED THE PLATE WAS INSERTED APPROPRIATELY AND FOR THE CORRECT INDICATION. BASED ON THE PERFORMED INVESTIGATION, THERE IS NO INDICATION FOR AN INCORRECTLY WORKING PRODUCT OR ANY DESIGN, MATERIAL OR MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PLATE WAS BROKEN AND A REVISION SURGERY WAS PERFORMED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE PLATE WAS BROKEN AND A REVISION SURGERY WAS PERFORMED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2836183 PRIMARY MANDIBULAR RECONSTRUCTION PLATE, 11-HOLES, WITH TEMPLATE IMPLANT JEY STRYKER LEIBINGER FREIBURG 55-15911 NA 04546540356376

Patients

Seq Age Sex Outcome Treatment
1 Unknown Required Intervention