FDA Adverse Event Malfunction Summary report: N

MICRODEBRIDER BLADES AND BURS

MDR report key: 14697119 · Received June 15, 2022

Report

Report Number
1045254-2022-00329
Event Type
Malfunction
Date Received
June 15, 2022
Date of Event
December 14, 2020
Report Date
June 15, 2022
Manufacturer
MEDTRONIC XOMED INC.
Product Code
EQJ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

PRODUCT ANALYSIS: VISUALLY, THE INNER SHAFT BROKE APPROXIMATELY 0.55 INCHES FROM THE DISTAL END OF THE INNER HUB. THE DISTAL OUTSIDE DIAMETER OF THE INNER HUB WAS WARPED AND DEFORMED WHICH WOULD HAVE RESULTED IN THE REPORTED EVENT. THE OUTSIDE DIAMETER OF THE INNER HUB SHOULD BE 0.330 ± 0.002 INCHES AND MEASURED 0.330 INCHES IN THE UNDAMAGED AREA AND UP TO 0.370 INCHES IN THE DEFORMED AREA WHICH WAS OUT OF SPECIFICATION. THE CHEVRONS ON THE PROXIMAL END OF THE HUB SHOWED WEAR AND HAD INDENTIONS. THE HUB BUSHING WAS WORN AND EXPOSED. THE HUB SEAL WAS STILL IN PLACE BUT WAS WORN. FUNCTIONAL TESTING COULD NOT BE PERFORMED DUE TO THE BROKEN STATE OF THE DEVICE. CONCLUSION: IN THE RETURNED CONDITION, THERE WAS AN OUT OF SPECIFICATION CONDITION THAT WAS RELATED TO THE COMPLAINT (DUE TO PHYSICAL DAMAGE) MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 0

A HEALTH CARE PROVIDER (HCP) REPORTED THAT A HANDPIECE WAS BROKEN PRIOR TO USE. THERE WAS NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1628033 MICRODEBRIDER BLADES AND BURS BUR, EAR, NOSE AND THROAT EQJ MEDTRONIC XOMED INC. XOM UNK M-BUR-BLD

Patients

Seq Age Sex Outcome Treatment
1 Unknown