FDA Adverse Event Death Summary report: N

PUMP MMT-1780KPK 670G PATHWAY BLACK MG

MDR report key: 14692976 · Received June 15, 2022

Report

Report Number
2032227-2022-213475
Event Type
Death
Date Received
June 15, 2022
Date of Event
January 1, 2022
Report Date
June 15, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OZP
UDI-DI
000000763000190460
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). THE CUSTOMER PASSED AWAY ON (B)(6) 2022. THE PUMP PASSED THE FUNCTIONAL TEST, INCLUDING THE SELF TEST, SLEEP CURRENT MEASUREMENT, ACTIVE CURRENT MEASUREMENT, REWIND TEST, PRIME/SEATING TEST, BASIC OCCLUSION TEST, OCCLUSION TEST, FORCE SENSOR TEST, DISPLACEMENT TEST AND THE DAT AT 0.0875 INCHES. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: MINOR SCRATCHED DISPLAY WINDOW, SCRATCHED CASE, PILLOWING KEYPAD OVERLAY, CRACKED KEYPAD OVERLAY AT THE SELECT BUTTON AND CRACKED RETAINER. THE TEST P-CAP AND RESERVOIR DOES LOCK IN PLACE IN THE RESERVOIR COMPARTMENT. HISTORY DOWNLOAD WAS SUCCESSFUL USING THUS AND CARELINK UPLOAD WAS SUCCESSFUL. THE PUMP DID NOT HAVE A BATTERY INSTALLED WHEN RECEIVED. PLEASE SEE BELOW FOR THE CUSTOMER'S DAILY TOTAL OF ALL INSULIN DELIVERED SURROUNDING THE DATE (B)(6) 2021 LISTED ON SMARTSOLVE AND THE DAYS PRIOR TO THAT DATE. THERE WAS NO DATA LISTED ON (B)(6) 2022. 12/25/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0, 12/26/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0, 12/27/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0, 12/28/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0, 12/29/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0, 12/30/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0, 12/31/2021 DAILY TOTAL OF ALL INSULIN DELIVERED: 0. THE PUMP PASSED THE FUNCTIONAL TESTING.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT CUSTOMER WAS PASSED AWAY. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1847568 PUMP MMT-1780KPK 670G PATHWAY BLACK MG ARTIFICIAL PANCREAS DEVICE SYSTEM, SINGLE HORMONAL CONTROL OZP MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1780KPK HG3MM2T 000000763000190460

Patients

Seq Age Sex Outcome Treatment
1 76 YR Female Death