T2SARS-COV-2 (EUA)
Report
- Report Number
- 3010097867-2022-00016
- Event Type
- Malfunction
- Date Received
- June 14, 2022
- Date of Event
- September 4, 2020
- Report Date
- June 14, 2022
- Manufacturer
- T2 BIOSYSTEMS, INC
- Product Code
- QJR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
GULF COAST MEDICAL CENTER REPORTED DISCREPANT ASSAY RESULTS DURING ASSAY VALIDATION - THE T2SARS-COV-2 PANEL (EUA), LOT WO-10199/EXPIRY 02/04/2021 RUN ON T2DX INSTRUMENT ISA1415000043 PRODUCED A NEGATIVE RESULT ON 04 SEP 2020, COMPARED TO THE POSITIVE SARS-COV-2 RESULT FROM THE CEPHEID GENEXPERT RUN ON (B)(6) 2020. THIS IS THE THIRD OF FIVE DISCREPANT NEGATIVE RESULTS OBTAINED DURING VALIDATION, WITH THE UNIQUE RUN ID: (B)(4). THE CUSTOMER RAN QC ON THE T2DX AND CONFIRMED INSTRUMENT PERFORMANCE. THE CEPHEID GENEXPERT ASSAY IS MORE SENSITIVE IN SARS-COV-2 DETECTION THAN THE T2SARS-COV-2 PANEL AND LIKELY CAUSE FOR THE DISCREPANT RESULT.
ON (B)(6) 2020, A CUSTOMER REPORTED THAT THE T2SARS-COV-2 PANEL REPORTED A NEGATIVE RESULT ON A KNOWN POSITIVE SAMPLE DURING SYSTEM VALIDATION. THE SAME SAMPLE REPORTED A POSITIVE RESULT ON THE CEPHEID GENEXPERT SYSTEM.