Narratives
W. L. GORE & ASSOCIATES, INC. (GORE) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY GORE, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. BLANK FIELDS PRESENT ON THIS REPORT INCLUDE REQUIRED FIELDS AND FIELDS DETERMINED TO BE NOT APPLICABLE. BLANK REQUIRED FIELDS INDICATE THAT THE INFORMATION WAS NOT PROVIDED, WAS DEEMED UNAVAILABLE OR WAS NOT APPLICABLE. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, GORE, OR ITS ASSOCIATES THAT THE DEVICE, GORE OR ITS ASSOCIATES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE A LEGAL ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY DEFECTS OR HAS MALFUNCTIONED, AS DEFINED FROM A LEGAL STANDPOINT. THESE WORDS ARE INCLUDED IN THE REPORT AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.
ON (B)(6) 2022, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. AFTER THE PARTIAL DEPLOYMENT OF THE TRUNK-IPSILATERAL LEG ENDOPROSTHESIS, IT WAS A DIFFICULT TO CANNULATE THE CONTRALATERAL GATE. THE PHYSICIAN USED THE CONSTRAINING SYSTEM TO REPOSITION THE DEVICE 1 CM DISTALLY. AFTER THE CANNULATION WAS SUCCESSFUL, THE DEVICE WAS REPOSITIONED AND MOVED PROXIMALLY ABOUT 1CM. AT THE FINAL ANGIOGRAPHY, AORTIC DISSECTION WAS OBSERVED AT THE PROXIMAL EDGE OF THE DEVICE. AN AORTIC EXTENDER COMPONENT WAS IMPLANTED TO CLOSE THE ENTRY. THE PATIENT TOLERATED THE PROCEDURE. REPORTEDLY THE DISSECTION WAS FORMED WHEN THE PHYSICIAN MOVED UP THE DEVICE PROXIMALLY WHILE REPOSITIONING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1788450 | GORE® EXCLUDER® AAA ENDOPROSTHESIS | SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | W. L. GORE & ASSOCIATES, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 87 YR | Male | Required Intervention |