BJORK-SHILEY 70 DEGREE CONVEXO-CONCAVE
Report
- Report Number
- 6000012-1998-00007
- Event Type
- Death
- Date Received
- January 30, 1998
- Date of Event
- October 10, 1988
- Report Date
- January 5, 1998
- Manufacturer
- SHILEY INC.
- Product Code
- LWQ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- ATTORNEY
Narratives
DEVICE EVALUATION SUMMARY: THE VALVE WAS EXAMINED WITH A LIGHT MICROSCOPE AT 20X MAGNIFICATION AND A FRACTOGRAPHIC ANALYSIS WAS DONE. THE FRACTOGRAPHIC ANALYSIS CONCLUDED THAT THE RIGHT LEG OF THE OUTLET STRUT FAILED FIRST. PROPAGATION WAS BY FATIGUE FOR BOTH THE RIGHT AND LEFT OUTLET STRUT LEGS. THERE WAS 70% BURNISHING ON THE LEFT OUTLET STRUT FACE. CRACK ORIGIN WAS ON THE INFLOW SIDE AT THE 6:00 POSITION FOR BOTH THE LEFT AND RIGHT OUTLET STRUT LEGS. THE ROOT CAUSE WAS DUE TO FATIGUE AT THE INITIATION SITES FOR BOTH LEGS. WEAR PATTERNS, SCRATCHES, AND INSTRUMENT MARKS TYPICAL FOR CLINICALLY EXPLANTED VALVES WERE OBSERVED ON THE STRUTS, THE FLANGE INNER DIAMETER, RIMS, AND DISC. THE FRACTOGRAPHIC FAILURE ANALYSIS INDICATED NO ANOMALOUS ATTRIBUTES WERE FOUND THAT MIGHT HAVE BEEN RESPONSIBLE FOR THE OUTLET STRUT FRACTURE.
ACCORDING TO THE INITIAL RPTR, THE PT'S BJORK-SHILEY CONVEXO-CONCAVE MITRAL VALVE WAS FOUND TO HAVE HAD AN OUTLET STRUT FRACTURE AFTER HIS BODY WAS EXHUMED AND AN AUTOPSY WAS PERFORMED. UPON RECEIPT OF THE VALVE AT SHILEY, IT WAS NOTED THAT THE VALVE WAS MISSING THE OUTLET STRUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | BJORK-SHILEY 70 DEGREE CONVEXO-CONCAVE | HEART VALVE | LWQ | SHILEY INC. | 25MBRCB | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Death | BSCC 70 DEGREE AORTIC HEART VALVE| (UNK TO 10/10/1988 (STOP'D)). |