BJORK-SHILEY 70 DEGREE CONVEXO-CONCAVE

MDR report key: 146809 · Received January 30, 1998

Report

Report Number
6000012-1998-00007
Event Type
Death
Date Received
January 30, 1998
Date of Event
October 10, 1988
Report Date
January 5, 1998
Manufacturer
SHILEY INC.
Product Code
LWQ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
ATTORNEY

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: THE VALVE WAS EXAMINED WITH A LIGHT MICROSCOPE AT 20X MAGNIFICATION AND A FRACTOGRAPHIC ANALYSIS WAS DONE. THE FRACTOGRAPHIC ANALYSIS CONCLUDED THAT THE RIGHT LEG OF THE OUTLET STRUT FAILED FIRST. PROPAGATION WAS BY FATIGUE FOR BOTH THE RIGHT AND LEFT OUTLET STRUT LEGS. THERE WAS 70% BURNISHING ON THE LEFT OUTLET STRUT FACE. CRACK ORIGIN WAS ON THE INFLOW SIDE AT THE 6:00 POSITION FOR BOTH THE LEFT AND RIGHT OUTLET STRUT LEGS. THE ROOT CAUSE WAS DUE TO FATIGUE AT THE INITIATION SITES FOR BOTH LEGS. WEAR PATTERNS, SCRATCHES, AND INSTRUMENT MARKS TYPICAL FOR CLINICALLY EXPLANTED VALVES WERE OBSERVED ON THE STRUTS, THE FLANGE INNER DIAMETER, RIMS, AND DISC. THE FRACTOGRAPHIC FAILURE ANALYSIS INDICATED NO ANOMALOUS ATTRIBUTES WERE FOUND THAT MIGHT HAVE BEEN RESPONSIBLE FOR THE OUTLET STRUT FRACTURE.

Description of Event or Problem · 1

ACCORDING TO THE INITIAL RPTR, THE PT'S BJORK-SHILEY CONVEXO-CONCAVE MITRAL VALVE WAS FOUND TO HAVE HAD AN OUTLET STRUT FRACTURE AFTER HIS BODY WAS EXHUMED AND AN AUTOPSY WAS PERFORMED. UPON RECEIPT OF THE VALVE AT SHILEY, IT WAS NOTED THAT THE VALVE WAS MISSING THE OUTLET STRUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BJORK-SHILEY 70 DEGREE CONVEXO-CONCAVE HEART VALVE LWQ SHILEY INC. 25MBRCB NA

Patients

Seq Age Sex Outcome Treatment
1 * Death BSCC 70 DEGREE AORTIC HEART VALVE| (UNK TO 10/10/1988 (STOP'D)).