FDA Adverse Event Malfunction Summary report: N

FRAGMENT SCREW, 2.5MMX20MM, HEXALOBE

MDR report key: 14674969 · Received June 13, 2022

Report

Report Number
1220246-2022-05065
Event Type
Malfunction
Date Received
June 13, 2022
Date of Event
May 23, 2022
Report Date
July 1, 2022
Manufacturer
ARTHREX, INC.
Product Code
HRS
UDI-DI
00888867027725
PMA / PMN Number
K123241
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

COMPLAINT NOT CONFIRMED. UPON VISUAL INSPECTION, IT WAS NOTED THAT THERE WERE SCRATCHES ALONG THE LENGTH OF THE SCREW. THERE WAS ALSO SLIGHT WEAR TO THE THREADS. THIS DAMAGE CAN MOST LIKELY BE CONTRIBUTED TO IMPLANTATION/EX-PLANTATION OF THE DEVICE. MEASUREMENT OF THE DEVICE SHOWS THAT THE DEVICES MEET DESIGN SPECIFICATIONS.

Description of Event or Problem · 0

ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2652CR CLAVICLE FRACTURE PLATE WAS IMPLANTED IN PATIENT ON (B)(6) 2022 DURING A CLAVICLE FRACTURE. PATIENT FELT THE PLATE BREAK AND ON (B)(6) 2022, IT WAS CONFIRM VIA X-RAYS THAT THE PLATE HAD BROKEN. A REVISION SURGERY HAS BEEN SCHEDULE BY THE SAME SURGEON THAT PERFORMED THE ORIGINAL PROCEDURE AND IT WILL TAKE PLACE AT THE SAME FACILITY. THIS REVISION SURGERY HAS BEEN SCHEDULED FOR (B)(6) 2022 TO REMOVE ALL THE IMPLANTED SCREWS ALONG WITH THE BROKEN PLATE. ADDITIONAL INFORMATION RECEIVED ON 5/26/2022: DURING ORIGINAL PROCEDURE, SURGEON IMPLANTED THE FOLLOWING: ONE AR-2652CR CLAVICLE FRACTURE PLATE, ONE AR-2665-16H FRAGMENT SCREW, ONE AR-2665-18H FRAGMENT SCREW, ONE AR-2665-20H FRAGMENT SCREW, ONE AR-2665-22H FRAGMENT SCREW, ONE AR-8835-16 LOW PROFILE SCREW, ONE AR-8835-18 LOW PROFILE SCREW, TWO AR-8835-20 LOW PROFILE SCREWS, ONE AR-8835L-16 LOW PROFILE LOCKING SCREW, AND THREE AR-8835L-18 LOW PROFILE LOCKING SCREW. ALL THESE IMPLANTS WERE REMOVED DURING REVISION SURGERY ON (B)(6) 2022. CASE WAS COMPLETED SUCCESSFULLY USING SYNTHES WITHOUT FURTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2813302 FRAGMENT SCREW, 2.5MMX20MM, HEXALOBE PLATE, FIXATION, BONE HRS ARTHREX, INC. FRAGMENT SCREW, 2.5MMX20MM, HEXALOBE 1103719 00888867027725

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other