FDA Adverse Event Malfunction Summary report: N

CLAV FRAC PLT,CENTRAL THIRD,RT,SS

MDR report key: 14674345 · Received June 13, 2022

Report

Report Number
1220246-2022-05055
Event Type
Malfunction
Date Received
June 13, 2022
Date of Event
May 23, 2022
Report Date
June 29, 2022
Manufacturer
ARTHREX, INC.
Product Code
HRS
UDI-DI
00888867027435
PMA / PMN Number
K143139
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

THE CONTRIBUTION OF THE DEVICE TO THE REPORTED EVENT COULD NOT BE DETERMINED AS THE DEVICE WAS NOT RETURNED FOR EVALUATION. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED FROM THE INFORMATION AVAILABLE AND WITHOUT DEVICE EVALUATION. IF THE DEVICE BECOMES AVAILABLE FOR EVALUATION, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

COMPLAINT CONFIRMED. UPON VISUAL INSPECTION IT WAS NOTED THAT THE PLATE WAS BROKEN WITHIN THE 5TH SCREW SLOT, BETWEEN THE CE MARK AND THE BATCH NUMBER. DENTS/SCRATCHES WERE NOTED ON THE TOP OF THE DEVICE AND WITHIN THE SCREW HOLES. THE THICKNESS AND WIDTH OF THE DEVICE WERE MEASURED AND FOUND TO BE WITHIN SPECIFICATIONS. THE CAUSE OF THIS REMAINS UNDETERMINED.

Description of Event or Problem · 0

ON (B)(6) 2022, IT WAS REPORTED BY A SALES REPRESENTATIVE VIA EMAIL THAT AN AR-2652CR CLAVICLE FRACTURE PLATE WAS IMPLANTED IN PATIENT ON (B)(6) 2022 DURING A CLAVICLE FRACTURE. PATIENT FELT THE PLATE BREAK AND ON (B)(6) 2022, IT WAS CONFIRM VIA X-RAYS THAT THE PLATE HAD BROKEN. A REVISION SURGERY HAS BEEN SCHEDULE BY THE SAME SURGEON THAT PERFORMED THE ORIGINAL PROCEDURE AND IT WILL TAKE PLACE AT THE SAME FACILITY. THIS REVISION SURGERY HAS BEEN SCHEDULED FOR (B)(6) 20222 TO REMOVE ALL THE IMPLANTED SCREWS ALONG WITH THE BROKEN PLATE. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022: DURING ORIGINAL PROCEDURE, SURGEON IMPLANTED THE FOLLOWING: ONEAR-2652CR CLAVICLE FRACTURE PLATE, ONE AR-2665-16H FRAGMENT SCREW, ONE AR-2665-18H FRAGMENT SCREW, ONE AR-2665-20H FRAGMENT SCREW, ONE AR-2665-22H FRAGMENT SCREW, ONE AR-8835-16 LOW PROFILE SCREW, ONE AR-8835-18 LOW PROFILE SCREW, TWO AR-8835-20 LOW PROFILE SCREWS, ONE AR-8835L-16 LOW PROFILE LOCKING SCREW, AND THREE AR-8835L-18 LOW PROFILE LOCKING SCREW. ALL THESE IMPLANTS WERE REMOVED DURING REVISION SURGERY ON (B)(6) 2022. CASE WAS COMPLETED SUCCESSFULLY USING SYNTHES WITHOUT FURTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1965566 CLAV FRAC PLT,CENTRAL THIRD,RT,SS PLATE, FIXATION, BONE HRS ARTHREX, INC. CLAV FRAC PLT,CENTRAL THIRD,RT,SS UNK 00888867027435

Patients

Seq Age Sex Outcome Treatment
1 Unknown Other