FDA Adverse Event Malfunction Summary report: N

BINAXNOW COVID-19 AG CARD

MDR report key: 14659566 · Received June 10, 2022

Report

Report Number
1221359-2022-02910
Event Type
Malfunction
Date Received
June 10, 2022
Date of Event
April 15, 2022
Report Date
August 11, 2022
Manufacturer
ABBOTT DIAGNOSTICS SCARBOROUGH, INC.
Product Code
QKP
UDI-DI
10811877011290
PMA / PMN Number
EUA202537
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REMAINDER OF THE INVESTIGATION REMAINS IN PROGRESS. A SUPPLEMENTAL REPORT WILL BE PROVIDED AFTER COMPLETION. RELATED MFR. REPORT NUMBERS: 1221359-2022-02895 THROUGH 1221359-2022-2930.

Additional Manufacturer Narrative · 0

INVESTIGATION REPORT: TESTING WAS PERFORMED AT ABBOTT DIAGNOSTICS SCARBOROUGH, INC. ON RETAINED KIT LOT 173244 WITH INTERNAL POSITIVE QUALITY CONTROL SAMPLES AND NEGATIVE QUALITY CONTROL SWABS. ALL TEST RESULTS WERE VALID AND PERFORMED AS EXPECTED. ADDITIONALLY, THE MANUFACTURING RECORDS AND QUALITY CONTROL RELEASE TESTING WAS REVIEWED FOR TEST KIT PART NUMBER 195-000 / LOT: 173244, TEST BASE PART NUMBER 195-430H / LOT: 170435. THE LOT MET THE REQUIRED RELEASE SPECIFICATIONS. A REVIEW OF THE COMPLAINTS REPORTED AS FALSE NEGATIVE PATIENT RESULTS (CONFIRMED AND UNCONFIRMED, CONFLICTING RESULTS) RELATED TO KIT LOT 173244 SHOWED THAT THE COMPLAINT RATE IS (B)(4). ABBOTT DIAGNOSTICS SCARBOROUGH, INC. WAS UNABLE TO DETERMINE THE EXACT ROOT CAUSE OF THE REPORTED ISSUE; HOWEVER, IT COULD BE RELATED TO THE SPECIFIC PATIENT SAMPLES.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THIRTY-SIX (36) FALSE NEGATIVE RESULTS WITH THE BINAXNOW COVID-19 AG CARD OCCURRING IN THE MONTH OF APR2022. THIS MFR. REPORT ADDRESSES TEST SIXTEEN (16) OF THIRTY-SIX (36) AND LOT: 173244 (TOTAL QUANTITY: 9). THE CUSTOMER REPORTED A FALSE NEGATIVE RESULT WITH THE BINAXNOW COVID-19 AG CARD PERFORMED ON (B)(6) 2022 ON A NASAL SAMPLE WITH THE KIT SWAB. PCR CONFIRMATION TESTING (PLATFORM: PERKIN-ELMER) GENERATED A POSITIVE RESULT AND WAS PERFORMED WITHIN 48 HOURS. PER THE CUSTOMER, THE CONFIRMATION TEST WAS PERFORMED WHEN EITHER SYMPTOMATIC PATIENTS RECEIVE NEGATIVE RESULTS WITH BINAXNOW COVID-19 AG CARD, OR UPON REQUEST OF THE PATIENT OR HEALTHCARE PROVIDER. THE PATIENT WAS ASYMPTOMATIC AND WAS NOT KNOWN TO BE COVID-19 POSITIVE AT THE TIME OF TESTING. THE CUSTOMER CONFIRMED THERE WAS NO PATIENT HARM DUE TO THE TEST RESULTS. ADDITIONALLY, THE CUSTOMER CONFIRMED THERE WAS NO DELAY OR IMPACT IN THE PATIENT'S TREATMENT. PER THE CUSTOMER, ROUTINE TESTING WAS BEING PERFORMED IN NON K-12 SCHOOLS AND NON LONG-TERM CARE FACILITIES AND THE AG CARDS HAVE BEEN FUNCTIONING AS EXPECTED. THE CUSTOMER DID NOT ALLEGE A MALFUNCTION OF THE DEVICE AND CONFIRMED THIS IS ROUTINE REPORTING PER THE BINAXNOW COVID-19 AG CARD INSTRUCTIONS FOR USE (IFU).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2290546 BINAXNOW COVID-19 AG CARD LATERAL FLOW IMMUNOASSAY IVD OF COVID-19 QKP ABBOTT DIAGNOSTICS SCARBOROUGH, INC. 173244 10811877011290

Patients

Seq Age Sex Outcome Treatment
1 75 YR Male