ISITE PACS
Report
- Report Number
- 2954704-2009-00007
- Event Type
- Malfunction
- Date Received
- May 27, 2009
- Date of Event
- May 1, 2009
- Report Date
- May 1, 2009
- Manufacturer
- PHILIPS HEALTHCARE INFORMATICS, INC.
- Product Code
- LLZ
- PMA / PMN Number
- K063267
- Removal / Correction Number
- Z-0118-2009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
ISITE PACS IS A SOFTWARE PRODUCT. WE ARE ABLE TO EVALUATE THE PRODUCT AT THE CUSTOMER SITE BY REMOTE ACCESS, AS WELL AS EVALUATE THE SAME PRODUCT VERSION IN-HOUSE. THIS PRODUCT VERSION IS CURRENT UNDER RECALL PER Z-0118-2009, (B)(4). THIS RECALL AFFECTS ISITE PACS VERSION 4.1.X - 4.1.45.2. WHEN DIFFERENT CR CASSETTE SIZES ARE USED WITH DIFFERENT VALUES PROVIDED BY THE MODALITIES IN PIXEL SPACING DICOM TAG (0028.0030), ISITE PACS DOES NOT USE THESE VALUES FOR MEASUREMENT CALCULATIONS. WHEN PERFORMING MEASUREMENTS IN ISITE PACS, THERE IS A POTENTIAL TO MISCALCULATE MEASUREMENTS UNDER THE FOLLOWING CONDITION: PIXEL SPACING (0028, 0030) AND IMAGER PIXEL SPACING (0018, 1164) DICOM TAG VALUES ARE BOTH PRESENT AND DIFFERENT. THERE ARE NO PROBLEMS WITH MANUALLY CALIBRATED IMAGES USING THE ISITE IMAGE CALIBRATION TOOL. LABELING INSTRUCTS CUSTOMERS TO MANUALLY CALIBRATE EACH PROJECTION X-RAY IMAGE USING A MARKER OF KNOWN SIZE PRESENT ON THE IMAGE ITSELF. THIS MARKER IS TO BE PLACED ON THE SAME PLANE, AS THE ANATOMY TO ENSURE THE MOST ACCURATE DISTANCE MEASUREMENTS. RECALL INITIATED TO ALIGN PRODUCT MEASUREMENT BEHAVIOR WITH CURRENT DICOM AND IHE MEASUREMENTS STANDARDS, BUT CUSTOMERS ARE STILL INSTRUCTED TO CALIBRATE/USE MARKER OF KNOWN SIZE FOR PROJECTION IMAGING. ORIGINAL MDR REPORT 2954704-2009-00006.
NO REPORT OF PATIENT HARM OR INJURY RELATED TO THIS EVENT. CUSTOMER REPORTED WHEN INSTALLING NEW MODALITY AND VALIDATING FOR MEASUREMENTS PER LABELING, CANNOT GET MEASUREMENTS TO WORK AS EXPECTED. REQUESTED INFORMATION REGARDING DICOM TAG USAGE FOR MEASUREMENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ISITE PACS | ISITE PACS | LLZ | PHILIPS HEALTHCARE INFORMATICS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |