THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S
Report
- Report Number
- 8010047-2022-09831
- Event Type
- Malfunction
- Date Received
- June 10, 2022
- Date of Event
- May 16, 2022
- Report Date
- August 29, 2022
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORP.
- Product Code
- GEI
- UDI-DI
- 04953170383557
- PMA / PMN Number
- K211838
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
- Health Professional
- Yes
Narratives
THE DEVICE IS NOT RETURNED, AS SUCH A DEFINITIVE ROOT CAUSE OF THE REPORTED COMPLAINT CANNOT BE DETERMINED AT THIS TIME. THIS EVENT IS UNDER INVESTIGATION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF THE INVESTIGATION OR UPON RECEIVING ADDITIONAL INFORMATION.
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE DEVICE EVALUATION AND THE LEGAL MANUFACTURER'S FINAL INVESTIGATION. PLEASE SEE THE UPDATES IN SECTIONS D9, H3, H4, H6, AND H10. THE DEVICE WAS RETURNED AND AN EVALUATION WAS COMPLETED FOR IT. DURING INSPECTION, OLYMPUS CONFIRMED THE REPORTED EVENT, THE PROBE WAS FOUND TO BE BROKEN. ADDITIONALLY, THE PROXIMAL SIDE OF THE TISSUE PAD WAS WORN AWAY; HOWEVER, THIS DEFECT IS NOT CONSIDERED SEVERE ENOUGH TO CAUSE A POTENTIAL ADVERSE EVENT. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. ALTHOUGH A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED, THE FOLLOWING STEP-BY-STEP SCENARIO LIKELY CAUSED THE EVENT: 1. SEAL & CUT¿ OUTPUT WAS ACTIVATED WHILE GRASPING A THICK TISSUE WITH THE TIP OF THE GRASPING SECTION. THE PROXIMAL SIDE OF THE TISSUE PAD COME IN CONTACT TO THE PROBE. AS A RESULT, THE PAD WAS WORN AWAY. 2. THE TISSUE PAD WAS WORN AWAY, CAUSING THE NON-INSULATED OF THE GRASPING SECTION TO TOUCH THE PROBE. 3. ¿SEAL & CUT¿ OUTPUT WAS ACTIVATED UNDER THIS SITUATION, THE SCRATCHES INDICATING THAT THE PROBE AND GRASPING SECTION WERE IN CONTACT WITH EACH OTHER WERE MADE. 4. THE PROBE RECEIVED AN OUTPUT LOAD IN SEAL & CUT MODE OR RECEIVED A LOAD WHEN GRASPING TISSUE. AS A RESULT, THE PROBE CRACKED FROM THE SCRATCH. 5. THE PROBE BROKE WHEN ADDED SOME LOAD. THIS FOLLOWING INFORMATION IS INCLUDED IN THE INSTRUCTIONS FOR USE (IFU): ¿DO NOT ACTIVATE OUTPUT IN SEAL & CUT MODE WHILE THE GRASPING SECTION IS CLOSED WITHOUT CONTACTING TISSUE OR VESSEL, OR ENSURING THAT TISSUE IS TRANSECTED. OTHERWISE, A LOCAL INCREASE OF THE TEMPERATURE DUE TO A FRICTION BETWEEN THE PROBE TIP AND THE GRASPING SECTION MAY RESULT IN VARIOUS FORMS OF DAMAGE IN THE PROBE TIP AND/OR THE TISSUE PAD, SUCH AS PREMATURE WEAR, BREAKAGE, DEFORMATION, AND/OR FALLING OFF INSIDE THE BODY CAVITY AND/OR PARTIAL SEPARATING.¿ ¿WHEN CUTTING AND VESSEL SEALING IS PERFORMED IN SEAL & CUT MODE, APPLY LIGHT TENSION ON THE TISSUE SO THAT USERS CAN CONFIRM IT IS TRANSECTED. ALSO, STOP ACTIVATION IMMEDIATELY AFTER TISSUE IS TRANSECTED. OTHERWISE, THE GRASPING SECTION, THE TISSUE PAD, OR THE PROBE TIP MAY BREAK AND FALL OFF, AND PARTIAL SEPARATING OF THE TISSUE PAD MAY OCCUR DUE TO A LOCAL INCREASE OF TEMPERATURE CAUSED BY THE FRICTION BETWEEN TISSUE PAD AND THE PROBE TIP DURING ACTIVATION.¿ ¿DURING THE TREATMENT, DO NOT ACTIVATE OUTPUT WHILE APPLYING THE PROBE TIP TO THE TISSUE WITH A STRONG FORCE, GRASPING THICK TISSUE, OR TWISTING THE HANDLE. ALSO, DO NOT INSERT THE HANDLE WHILE THE HANDLE IS TWISTED WITH RESPECT TO THE TISSUE, DO NOT GRASP IT, AND DO NOT ACTIVATE THE OUTPUT. OTHERWISE, THE PROBE TIP AND/OR GRASPING SECTION MAY BE DAMAGED, WHICH MAY RESULT IN FALLING OF THE PROBE TIP AND/OR TISSUE PAD.¿ OLYMPUS WILL CONTINUE TO MONITOR THE PERFORMANCE OF THIS DEVICE.
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, DURING A THERAPEUTIC HYSTERECTOMY PROCEDURE A U504 PROBE DAMAGE ERROR MESSAGE WAS RECEIVED FOR THE DEVICE. THE DEVICE WAS BEING USED IN A REGULAR MANNER. SUBSEQUENTLY, THE DEVICE PROBE BROKE OFF AND FELL INTO THE PATIENT BODY. THE BROKEN PIECE WAS RETRIEVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITH A NEW SIMILAR DEVICE WITH A DELAY OF FIVE MINUTES. THE INTELLIGENT TISSUE MONITORING (ITM) FUNCTIONALITY WAS OFF AS IS USUAL. USER UNDERSTANDS THE FUNCTIONALITY OF ITM, WHICH IS THAT ITM IS A SUPPORT FUNCTION, AND THERE IS A POSSIBILITY THAT ITM CANNOT DETECT THE TISSUE ALREADY CUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2449311 | THUNDERBEAT 5 MM, 35 CM, FRONT-ACTUATED GRIP TYPE S | ULTRASONIC SURGICAL DEVICE | GEI | OLYMPUS MEDICAL SYSTEMS CORP. | TB-0535FCS | KR185564 | 04953170383557 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | USG-400 VERSION 2. |