FDA Adverse Event Malfunction Summary report: N

BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE

MDR report key: 14653125 · Received June 10, 2022

Report

Report Number
3003152976-2022-00257
Event Type
Malfunction
Date Received
June 10, 2022
Date of Event
May 13, 2022
Report Date
June 15, 2022
Manufacturer
BECTON DICKINSON, S.A.
Product Code
FMF
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: MULTIPLE PHOTOS WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. THROUGH VISUAL INSPECTION, SILICONE IS VISIBLE INSIDE THE SYRINGE. A DEVICE HISTORY REVIEW WAS PERFORMED FOR REPORTED LOT 2202013, NO DEVIATIONS OR NON-CONFORMANCES WERE IDENTIFIED DURING THE MANUFACTURING PROCESS THAT COULD HAVE CONTRIBUTED TO THIS ISSUE. LUBRICANT IS EMPLOYED DURING THE SYRINGE ASSEMBLY PROCESS TO LUBRICATE THE CYLINDERS IN THE SILICONE STATION. THE SILICONE EMPLOYED IN THIS PRODUCT IS A MEDICAL GRADE SILICONE AUTHORIZED FOR PRODUCT USE. SILICONE CONTENT TESTS ARE PERFORMED DURING THE MANUFACTURING PROCESS OF EACH LOT NUMBER. RESULTS WERE REVIEWED FOR LOT 2202013 AND FOUND TO BE WITHIN SPECIFICATION. WHILE WE CANNOT IDENTIFY A DIRECT ISSUE, SILICONE CAN BE VISIBLE DUE TO ISSUES RELATED TO POOR DISTRIBUTION INSIDE THE BARREL. H3 OTHER TEXT : SEE H10.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND WITHIN THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DISCOVERED AN OILY LIQUID IN THE SYRINGE WHEN OPENED THE OUTER PACKAGE.

Description of Event or Problem · 0

IT WAS REPORTED THAT FOREIGN MATTER WAS FOUND WITHIN THE BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DISCOVERED AN OILY LIQUID IN THE SYRINGE WHEN OPENED THE OUTER PACKAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
631432 BD PLASTIPAK¿ CONCENTRIC LUER LOCK SYRINGE PISTON SYRINGE FMF BECTON DICKINSON, S.A. 2202013

Patients

Seq Age Sex Outcome Treatment
1 Unknown