FDA Adverse Event Malfunction Summary report: N

10NM TORQUE WRENCH 11MM ACROSS THE FLATS

MDR report key: 14643335 · Received June 9, 2022

Report

Report Number
2939274-2022-02151
Event Type
Malfunction
Date Received
June 9, 2022
Report Date
June 9, 2022
Manufacturer
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
Product Code
HXC
UDI-DI
10705034772233
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). ADDITIONAL NARRATIVE: COMPLAINANT PART IS EXPECTED TO BE RETURNED FOR MANUFACTURER REVIEW/INVESTIGATION BUT HAS YET TO BE RECEIVED. REPORTER IS A J&J EMPLOYEE. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED. PRODUCT WAS NOT RETURNED. BASED ON THE INFORMATION AVAILABLE, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THIS COMPLAINT WILL BE ACCOUNTED FOR AND MONITORED VIA POST MARKET SURVEILLANCE ACTIVITIES. IF ADDITIONAL INFORMATION IS MADE AVAILABLE, THE INVESTIGATION WILL BE UPDATED AS APPLICABLE. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H6: PART # 388.261. SYNTHES LOT # H812484-05. SUPPLIER LOT # H812484-05. RELEASE TO WAREHOUSE DATE: 27 JUNE 2019. SUPPLIER: (B)(6). NO NCRS WERE GENERATED DURING PRODUCTION. REVIEW OF THE DEVICE HISTORY RECORD(S) SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THIS PRODUCT, AND ANY SUB-COMPONENTS, WHICH WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION. THE REPAIR TECHNICIAN REPORTED THAT THE DEVICE HAD FAILED CALIBRATION AND FAILED TORQUE TEST IN COUNTERCLOCKWISE DIRECTION. TORQUE TEST HIGH IS THE REASON FOR REPAIR. THE ITEM IS NOT REPAIRABLE PER THE INSPECTION SHEET. THE CAUSE OF THE ISSUE IS TORQUE TEST FAILED HIGH. THE EVALUATION WAS CONFIRMED. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H10: ADDITIONAL NARRATIVE: DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE. THE DEVICE WAS RECEIVED, THE INVESTIGATION COULD NOT BE COMPLETED, AND NO CONCLUSION COULD BE DRAWN, AS PRODUCT IS ENTERING THE COMPLAINT SYSTEM.

Description of Event or Problem · 0

DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING EVALUATION,10NM TORQUE WRENCH 11MM ACROSS THE FLATS HAD FAILED CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE (1) DEVICE 10NM TORQUE WRENCH 11MM ACROSS THE FLATS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
893709 10NM TORQUE WRENCH 11MM ACROSS THE FLATS WRENCH HXC WRIGHTS LANE SYNTHES USA PRODUCTS LLC 388.261 H812484-05 10705034772233

Patients

Seq Age Sex Outcome Treatment
1 Unknown