FDA Adverse Event Malfunction Summary report: N

PENTAX

MDR report key: 14641187 · Received June 9, 2022

Report

Report Number
9610877-2022-54484
Event Type
Malfunction
Date Received
June 9, 2022
Date of Event
May 18, 2022
Report Date
June 9, 2022
Manufacturer
HOYA CORPORATION PENTAX TOKYO OFFICE
Product Code
EOB
UDI-DI
04961333070589
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL FNL-10RP3 IS AVAILABLE IN THE USA WITH A 510K NUMBER K951196.WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE IMAGE GUIDE FIBER BUNDLE (CFB) WAS BROKEN. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE EXCESSIVE FORCE APPLIED ON THE IMAGE GUIDE FIBER BUNDLE (CFB).BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.

Description of Event or Problem · 0

THE TIME OF EVENT IS UNKNOWN.THERE WAS NO REPORT OF PATIENT HARM.FIBER IMAGE FAILURE(BROKEN).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2420438 PENTAX FIBER NASO PHARYNGO LARYNGOSCOPE EOB HOYA CORPORATION PENTAX TOKYO OFFICE FNL-10RBS 04961333070589

Patients

Seq Age Sex Outcome Treatment
1 Unknown