FDA Adverse Event
Malfunction
Summary report: N
PENTAX
MDR report key: 14641113
·
Received June 9, 2022
Report
- Report Number
- 9610877-2022-54429
- Event Type
- Malfunction
- Date Received
- June 9, 2022
- Date of Event
- May 23, 2022
- Report Date
- June 9, 2022
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- FDS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
THIS DEVICE IS CLASSIFIED AS IMPORT FOR EXPORT, THEREFORE 510K IS NOT APPLICABLE. MODEL EG-1690K IS AVAILABLE IN THE USA WITH A 510K NUMBER K131902. WE CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE AIR/WATER CHANNEL CLOGGED. BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE INSUFFICIENT REPROCESSING AT THE FACILITY ON THE AIR/WATER CHANNEL. BASED ON THE TECHNICAL REPORT "HR-RPT-0630(AIR/WATER & JET WATER CHANNELS)" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
Description of Event or Problem · 0
THE TIME OF EVENT IS UNKNOWN.THERE WAS NO REPORT OF PATIENT HARM.AIR/WATER TUBE CLOGGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2449621 | PENTAX | VIDEO GASTROSCOPE | FDS | HOYA CORPORATION PENTAX TOKYO OFFICE | EG16-K10 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |