FDA Adverse Event
Injury
Summary report: N
VARITRONICS
MDR report key: 146408
·
Received January 30, 1998
Report
- Report Number
- 146408
- Event Type
- Injury
- Date Received
- January 30, 1998
- Date of Event
- December 3, 1997
- Report Date
- December 19, 1997
- Manufacturer
- VARITRONICS
- Product Code
- HMT
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
DURING AN EYELID OPERATION: THE ELECTROSURGICAL OUTPUT SETTING WAS INCREASED FROM A 10 WATT SETTING TO 15 AND THEN TO 20 WATTS IN ORDER TO OBTAIN A STRONGER SPARK TO EASE THE CUTTING ACTION. WHEN THE PENCIL WAS ACTIVATED ON THE 20 WATT SETTING, A FLASH WAS EMITTED FROM THE TIP. THE CARDBOARD "NEVYAS" RETRACTOR IGNITED AND SPREAD TO THE PLASTIC DRAPE. THE FIRE WAS IMMEDIATELY EXTINGUISHED BY THE SURGEON. OXYGEN WAS TURNED OFF BY THE ANESTHESIOLOGIST. BURNED AREAS ON THE PT'S FACE WERE OBSERVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VARITRONICS | DRAPE RETRACTOR | HMT | VARITRONICS | NEVYAS | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Hospitalization| R |