FDA Adverse Event
Malfunction
Summary report: N
ESPRIT VENTILATOR
MDR report key: 1463893
·
Received September 8, 2009
Report
- Report Number
- 2031642-2009-00202
- Event Type
- Malfunction
- Date Received
- September 8, 2009
- Date of Event
- August 10, 2009
- Report Date
- August 10, 2009
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K981072
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE VENTILATOR ALARMED AND RESTARTED WHILE IN USE ON A PATIENT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM. THE MANUFACTURER'S SERVICE TECHNICIAN WAS ABLE TO DUPLICATE THE CUSTOMER REPORTED PROBLEM. THE SERVICE TECHNICIAN REPLACED THE MMI PCB BOARD TO CORRECT THE FINDINGS. EXTENDED SELF TESTING (EST) AND PERFORMANCE VERIFICATION TESTING WAS COMPLETED, AND ALL TESTS PASSED PER OPERATING SPECIFICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |