FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 1463893 · Received September 8, 2009

Report

Report Number
2031642-2009-00202
Event Type
Malfunction
Date Received
September 8, 2009
Date of Event
August 10, 2009
Report Date
August 10, 2009
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR ALARMED AND RESTARTED WHILE IN USE ON A PATIENT. THE CUSTOMER REPORTED THERE WAS NO PATIENT HARM. THE MANUFACTURER'S SERVICE TECHNICIAN WAS ABLE TO DUPLICATE THE CUSTOMER REPORTED PROBLEM. THE SERVICE TECHNICIAN REPLACED THE MMI PCB BOARD TO CORRECT THE FINDINGS. EXTENDED SELF TESTING (EST) AND PERFORMANCE VERIFICATION TESTING WAS COMPLETED, AND ALL TESTS PASSED PER OPERATING SPECIFICATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1