UNKNOWN AVENIR STEM
Report
- Report Number
- 0001822565-2022-01741
- Event Type
- Injury
- Date Received
- June 8, 2022
- Date of Event
- February 27, 2022
- Report Date
- August 5, 2022
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KWA
- PMA / PMN Number
- NI
- Removal / Correction Number
- NI
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). REPORT SOURCE: HARTFORD J.M., GRAW B.P, FROSCH D.L. (2022) REDUCED INCIDENCE OF PERIOPERATIVE PERIPROSTHETIC FRACTURES USING HYBRID RASP-IMPACTION BROACHING OVER IMPACTION BROACHING WHEN USING THE DIRECT ANTERIOR APPROACH FOR TOTAL HIP ARTHROPLASTY. ARTHROPLASTY TODAY VOL. 15 (2022) 75-80. HTTPS://WWW.ARTHROPLASTYTODAY.ORG/ARTICLE/S2352-3441(22)00069-3/FULLTEXT. CUSTOMER HAS INDICATED THAT THE PRODUCT WILL NOT BE RETURNED TO ZIMMER BIOMET FOR INVESTIGATION, PRODUCT LOCATION UNKNOWN. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE INVESTIGATION IS IN PROCESS. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. ADDITIONAL INFORMATION DOES NOT CHANGE THE ROOT CAUSE OF THE PREVIOUS INVESTIGATION. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. COMPONENT CODE: MECHANICAL - (G04) STEM. REPORTED EVENT WAS UNABLE TO BE CONFIRMED. DEVICE HISTORY RECORD (DHR) REVIEW WAS UNABLE TO BE PERFORMED AS THE ITEM AND LOT NUMBER OF THE DEVICE INVOLVED IN THE EVENT IS UNKNOWN. ROOT CAUSE COULD NOT BE DETERMINED AS THE NECESSARY INFORMATION TO ADEQUATELY INVESTIGATE THE REPORTED EVENT WAS NOT PROVIDED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED PATIENT EXPERIENCED AN INTRAOPERATIVE FRACTURE OF THE CALCAR OCCURRED DURING IMPLANT INSERTION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
IT WAS REPORTED PATIENT EXPERIENCED AN INTRAOPERATIVE FRACTURE OF THE CALCAR OCCURRED DURING IMPLANT INSERTION. THE SURGEON USED CERCLAGE WIRE TO ADDRESS THE FRACTURE. SURGEON BELIEVED EXCESSIVE HOOP FORCES AND OSTEOPOROSIS WAS THE CAUSE OF THE FRACTURE. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
NO ADDITIONAL INFORMATION ON THE REPORTED EVENT.
NO ADDITIONAL INFORMATION ON THE REPORTED EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 18296 | UNKNOWN AVENIR STEM | PROSTHESIS, HIP | KWA | ZIMMER BIOMET, INC. | N/A | NI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Other |