ACTIVA PC
Report
- Report Number
- 3004209178-2009-04785
- Event Type
- Malfunction
- Date Received
- July 13, 2009
- Date of Event
- June 16, 2009
- Report Date
- June 16, 2009
- Manufacturer
- MPROC, JUNCOS
- Product Code
- MHY
- PMA / PMN Number
- 96009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
(B) (4).
IT WAS REPORTED THAT FOLLOWING IMPLANT SURGERY, HIGH IMPEDANCES WERE REPORTED. ELECTRODE SHOWS 'OUT OF RANGE' BOX ON THE 8840 AT 3V/80PW/100HZ ON THE RIGHT SIDE WAS ONLY C11 AT 2164. READINGS ON THE LEFT SIDE AT THE SAME SETTINGS C0: 3161, C1: 5289, C2: 11739, C3: 2354, 01: 7155, 02: 1582003: 5432, 12: 14372, 13: 7288, 23: 13675. FIVE MINUTES LATER, LEFT SIDE IMPEDANCES WERE TESTED AGAIN AND NOTED: C0: 12688 OHMS, C1: 7288 OHMS, C2: 35409 OHMS, C3: 2778 OHMS, 01: 15431 OHMS, 02:>40,000 OHMS, 03: 15431 OHMS, 12: 3297 OHMS, 13: 9920 OHMS, 23: 36223 OHMS. THE PT WAS SEEN AGAIN THE NEXT DAY ((B) (6) 2009) AND IMPEDANCES WERE CHECKED AGAIN. THE IMPEDANCE MEASUREMENTS ON THE RIGHT SIDE WERE ALL NORMAL EXCEPT C11 AT 2,369. THE LEFT SIDE IMPEDANCES MEASURED C & 0 15,584; C & 1 8,518; C & 2 36,223; C & 3 3,218; 0 & 1 18,308; 0 & 2 >40,000; 0 & 3 18,857; 1 & 2 30,297; 1 & 3 11,115; 2 & 3 37,076. IT WAS FELT THERE WAS LIKELY AN OPEN CIRCUIT WITH THE #2 CONTACT ON THE LEFT SIDE AND THAT THE OTHER CONTACTS WERE OK. THE PT WAS SEEN AGAIN APPROX TWO WEEKS LATER ((B) (6) 2009). LEFT SIDE IMPEDANCES WERE THE SAME AS THOSE MEASURED ON (B) (6) 2009. FURTHER HIGH IMPEDANCES WERE SEEN ON THE RIGHT SIDE: C & 11 2369, 8 & 9 2423, 8 & 10 2651, 8 & 11 3244, 9 & 10 3085, 9 & 11 3908, AND 10 & 11 3595. THE PT WAS BEING SEEN FOR PROGRAMMING ON (B) (6) 2009. NO PT SYMPTOMS OR OUTCOME WAS REPORTED. ADD'L INFO HAS BEEN REQUESTED, BUT WAS NOT AVAILABLE AS OF THE DATE OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACTIVA PC | MHY | MPROC, JUNCOS | 37601 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | EXPLANTED:| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL UNK, LOT# UNK| PROGRAMMER: MODEL 8840, LOT# UNK| EXPLANTED:| IMPLANTED:| EXTENSION: MODEL UNK, LOT# UNK| EXPLANTED:| IMPLANTED:| LEAD: MODEL UNK, LOT# UNK| LEAD: MODEL UNK, LOT# UNK| IMPLANTED: |