OXFORD PKS ANATOMIC ARCOM MENISCAL BRG - XSMALL RIGHT SIZE 4
Report
- Report Number
- 3002806535-2022-00252
- Event Type
- Injury
- Date Received
- June 8, 2022
- Date of Event
- May 6, 2022
- Report Date
- September 20, 2022
- Manufacturer
- BIOMET UK LTD.
- Product Code
- NRA
- UDI-DI
- 05019279073023
- PMA / PMN Number
- P010014
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4) CONCOMITANT MEDICAL DEVICES: OXF PH3 CEMENTLESS FEM SZ XSM; ITEM# 154912; LOT# 7055221, OXF UNI TIB TRAY SZ B RM PMA; ITEM# 154721; LOT# 7083494. REPORTV SOURCE FOREIGN: (B)(6). MULTIPLE MDR REPORTS WERE FILED FOR THIS EVENT, PLEASE SEE ASSOCIATED REPORTS: 3002806535-2022-00249, 3002806535-2022-00251. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS; HOWEVER, AN INVESTIGATION OF THE REPORTED EVENT IS IN PROGRESS. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL MEDWATCH 3500A WILL BE SUBMITTED.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. D10: MEDICAL DEVICES: OXF PH3 CEMENTLESS FEM SZ XSM; ITEM# 154912; LOT# 7055221. OXF UNI TIB TRAY SZ B RM PMA; ITEM# 154721; LOT# 7083494. OXF TRL BRG W/SLOTS XSM 4MM; ITEM# 32-422694; LOT# ZB130301. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4, B5, D9, G3, G6, H1, H2, H3, H6, H10. DIMENSIONAL ANALYSIS DETERMINED THAT THE PRODUCT, WHERE MEASURED, WAS OVERSIZE TO PRINT SPECIFICATIONS FOR DIMENSIONS A, B AND C. THIS IS AS EXPECTED BECAUSE THE BEARING HAS LIKELY BEEN AUTOCLAVED STERILISED AT THE HOSPITAL BEFORE RETURN AND ALSO HAS BEEN ADDITIONALLY SOAKED STERILISED FOR 5.5 HOURS UPON RETURN. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING. DEVICES ARE USED FOR TREATMENT. MEDICAL RECORDS WERE NOT PROVIDED. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT DURING AN INITIAL KNEE ARTHROPLASTY THE BEARING SPUN OUT AND FELL OFF TO THE POSTERIOR JOINT CAPSULE SIDE. THE SURGEON THEN DECIDED TO USE A DIFFERENT SYSTEM, AND THE FEMUR AND TIBIA COMPONENTS WERE REMOVED. DURING REMOVAL, THE PATIENT¿S FEMUR BONE CRACKED. SURGERY WAS DELAYED FOR ABOUT 40 MINUTES. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2438337 | OXFORD PKS ANATOMIC ARCOM MENISCAL BRG - XSMALL RIGHT SIZE 4 | OXFORD PARTIAL KNEE SYSTEM | NRA | BIOMET UK LTD. | N/A | 7029590 | 05019279073023 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Hospitalization| R | SEE H10 NARRATIVE| SEE H10 NARRATIVE. |