FDA Adverse Event Malfunction Summary report: N

640G INSULIN PUMP MMT-1711K

MDR report key: 14629502 · Received June 8, 2022

Report

Report Number
2032227-2022-205838
Event Type
Malfunction
Date Received
June 8, 2022
Date of Event
November 6, 2021
Report Date
June 7, 2022
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
UDI-DI
000000763000317089
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LO
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

(B)(4). S/W VER. 4.11D. RETAINER RING = BLACK. CUSTOMER RETURNED PUMP FOR AN ALLEGED PUMP ERROR 4, 14 AND 42 ON (B)(6) 2021. TEST P-CAP AND RESERVOIR LOCKED PROPERLY INTO RESERVOIR COMPARTMENT DURING TESTING. THE PUMP PASSED THE DISPLACEMENT TEST, REWIND TEST AND SELF TEST. NO UNEXPECTED PUMP ERROR 4, 14 AND 42 ALARMS DURING TESTING HOWEVER, PUMP ERROR 4 ALARM (LINENUMBER = 2727: 401:457 ; FILENUMBER = 32122: 38: 13 ) IS FOUND IN THE FORMATTED HISTORY FILE ON (B)(6) 2021 23:56:11.000 DUE TO A POSSIBLE HARDWARE ISSUE. SUCCESSFULLY DOWNLOADED HISTORY FILES AND TRACES USING THUS SOFTWARE. PUMP WAS CUT OPEN TO PERFORM VISUAL INSPECTION AND FOUND NO EVIDENCE OF PHYSICAL OR MOISTURE DAMAGE ON THE PCBA1, PCBA2, FORCE SENSOR, MOTOR AND VIBRATOR ASSEMBLY NOTED. THE FOLLOWING WERE NOTED DURING VISUAL INSPECTION: SCRATCHED CASE, CRACKED CASE (CORNER OF BELT CLIP RAILS) AND CRACKED BATTERY TUBE THREADS.¿ IN SUMMARY, CUSTOMER ALLEGED PUMP ERROR 4, 14 AND 42 CONFIRMED. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE 640G INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES. MEDTRONIC, INC. (MEDTRONIC) IS SUBMITTING THIS REPORT TO COMPLY WITH 21 C.F.R. PART 803, THE MEDICAL DEVICE REPORTING REGULATION. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION IN THE TIME ALLOTTED AND HAS PROVIDED AS MUCH INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY "DEFECTS" OR HAS "MALFUNCTIONED". THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA, TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REPORTING PURSUANT TO PART 803. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE IT BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT.

Description of Event or Problem · 0

INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MULTIPLE ERROR ALARMS. CUSTOMER WAS ABLE TO SUCCESSFULLY CLEAR THE ALARM AND REWIND INSULIN PUMP. CUSTOMER STATED THAT THE INSULIN PUMP WAS PASSED IN SELF TEST. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
825860 640G INSULIN PUMP MMT-1711K PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-1711K 000000763000317089

Patients

Seq Age Sex Outcome Treatment
1 Unknown