PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM
Report
- Report Number
- 2024168-2022-06047
- Event Type
- Injury
- Date Received
- June 7, 2022
- Date of Event
- May 18, 2022
- Report Date
- August 9, 2022
- Manufacturer
- ABBOTT VASCULAR
- Product Code
- MGB
- PMA / PMN Number
- P960043
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
(B)(4). THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED CUFF MISS WAS NOT CONFIRMED AS NOT ALL COMPONENTS WERE RETURNED FOR ANALYSIS. A REVIEW OF THE MANUFACTURING RECORDS IDENTIFIED NO MANUFACTURING NONCONFORMITIES ISSUED TO THE REPORTED LOT THAT WOULD HAVE CONTRIBUTED TO THIS EVENT. THE PROGLIDE DEVICE WAS USED IN THE ANTERIOR TIBIAL ARTERY. IT SHOULD BE NOTED THAT THE ELECTRONIC PROGLIDE INSTRUCTIONS FOR USE (IFU), STATES: THE PERCLOSE PROGLIDE SUTURE MEDIATED CLOSURE SYSTEM IS INDICATED FOR THE PERCUTANEOUS DELIVERY OF SUTURE FOR CLOSING THE COMMON FEMORAL ARTERY AND VEIN ACCESS SITE OF PATIENTS WHO HAVE UNDERGONE DIAGNOSTIC OR INTERVENTIONAL CATHETERIZATION PROCEDURES. IT IS UNKNOWN IF THE IFU DEVIATION CONTRIBUTED TO THE REPORTED DIFFICULTIES. THE REPORTED DIFFICULTY AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4 - LOT NO UPDATED FROM 1062641 TO 1030242. D4 - EXPIRATION DATE UPDATED FROM 4/30/2023 TO 1/31/2023 H4 - DEVICE MFG DATE UPDATED FROM 6/26/2021 TO 3/2/2021
IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF THE LEFT ANTERIOR TIBIAL ARTERY WAS ATTEMPTED WITH A PROGLIDE DEVICE AFTER A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY INTERVENTIONAL PROCEDURE USING A 8F SHEATH. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED. MANUAL COMPRESSION WAS USED TO ACHIEVE HEMOSTASIS. THERE WAS NO ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1141305 | PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM | SUTURE MEDIATED CLOSURE | MGB | ABBOTT VASCULAR | 1030242 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |