FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN BLOOD TUBES

MDR report key: 14607800 · Received June 6, 2022

Report

Report Number
1917413-2022-00305
Event Type
Malfunction
Date Received
June 6, 2022
Date of Event
May 20, 2022
Report Date
May 25, 2022
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
50382903679626
PMA / PMN Number
K945952
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. INVESTIGATION SUMMARY: MAT: 367962. LOT: 2018937. NO CUSTOMER SAMPLES AND NO PHOTOS WERE RECEIVED IN SUPPORT OF THIS COMPLAINT. THEREFORE, 10 PRODUCTION LOT IN-HOUSE RETENTION TUBES FROM THE BD INVENTORY WERE INSPECTED AND FUNCTIONALLY TESTED AND THE CUSTOMER'S INDICATED FAILURE MODE OF UNDERFILL WAS NOT OBSERVED. BD WAS UNABLE TO CONFIRM THE CUSTOMER¿S INDICATED FAILURE MODE NO SAMPLES WERE RECEIVED, AND THE DEFECT WAS NOT OBSERVED IN THE RETENTION SAMPLE TESTING. THE DEVICE HISTORY RECORDS WERE REVIEWED WITH NO ISSUES BEING IDENTIFIED. THERE WERE NO RELATED QUALITY NOTIFICATIONS. ALL PROCESS AND FINAL INSPECTIONS COMPLY WITH SPECIFICATION REQUIREMENTS. THE QUANTITY OF BLOOD DRAWN INTO EVALUATED TUBES VARIES WITH ALTITUDE, AMBIENT TEMPERATURE, BAROMETRIC PRESSURE, TUBE AGE, VENOUS PRESSURE, AND FILLING TECHNIQUE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED WHEN USING THE BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN N (LH) 83 UNITS BLOOD COLLECTION TUBES HAVE NO VACUUM THUS RESULTING IN DRAW VOLUME OF TUBES. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. THE CUSTOMER STATED: "MATERIAL NO. 367962 BATCH NO. 2018937. IT WAS REPORTED BY THE CUSTOMER THAT TUBES HAVE NO VACUUM THUS RESULTING IN DRAW VOLUME OF TUBES."

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1074369 BD VACUTAINER® PST¿ GEL AND LITHIUM HEPARIN BLOOD TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 367962 2018937 50382903679626

Patients

Seq Age Sex Outcome Treatment
1 Unknown