PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN)
Report
- Report Number
- 1419937-2022-00052
- Event Type
- Injury
- Date Received
- June 6, 2022
- Date of Event
- May 8, 2022
- Report Date
- June 6, 2022
- Manufacturer
- MEDELA LLC
- Product Code
- HGX
- UDI-DI
- 00020451401447
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
THE CUSTOMER WAS SENT A REPLACEMENT PUMP AND RETURN OF HER ORIGINAL PUMP WAS REQUESTED FOR TESTING/EVALUATION. IN FOLLOW UP WITH A COMPLAINT HANDLER ON (B)(6) 2022, THE CUSTOMER INDICATED THAT SHE DEVELOPED MASTITIS ON (B)(6) 2022 AND WAS PRESCRIBED AN ANTIBIOTIC. SHE INDICATED THAT THE REPLACEMENT PUMP WAS WORKING WITHOUT ISSUE AND HER MASTITIS IS BETTER. THE DEVICE WAS RETURNED AND EVALUATED ON (B)(6) 2022 AND IT WAS DETERMINED THAT THE DC JACK WAS MISALIGNED AND THE PUMP HOUSING, KEYPAD, CHASSIS AND PC BOARD ARE DAMAGED. THE CUSTOMER'S REPORT OF BROKEN DAMAGED DC JACK WAS CONFIRMED. BASED ON THE RESULTS OF OUR INTERNAL INVESTIGATION ((B)(4)), IT CANNOT BE DEFINITIVELY CONCLUDED THAT THE PUMP CAUSED OR CONTRIBUTED TO THE CUSTOMER¿S MASTITIS. THE ESTIMATED INCIDENCE OF MASTITIS IN LACTATING WOMEN, WHETHER USING A BREAST PUMP OR NOT, ACCORDING TO PUBLISHED CLINICAL LITERATURE CAN BE AS HIGH AS 33%. IN FACT, CLINICAL GUIDELINES SUGGEST THE USE OF A BREAST PUMP TO FACILITATE WITHDRAWAL OF BREAST MILK DURING BOUTS OF MASTITIS. THE COMPLAINT RATE OF MASTITIS ACROSS ALL REPORTED FAILURES, ACROSS ALL MEDELA BREAST PUMPS, IS 0.008% FOR THE PERIOD OF JANUARY 2013 TO AUGUST 2017. MASTITIS IS USUALLY A BENIGN, SELF-LIMITING INFECTION WITH FEW CONSEQUENCES FOR THE SUCKLING INFANT. THE RISK OF MASTITIS IS HIGHER AMONG WOMEN WHO HAVE BREASTFED PREVIOUSLY, ESPECIALLY THOSE WITH A HISTORY OF MASTITIS. RIORDAN & WAMBACH, 4TH ED. P. 294: BREASTFEEDING AND HUMAN LACTATION. HOWEVER IN THIS CASE, THE MOTHER¿S MASTITIS REQUIRED PROMPT MEDICAL ATTENTION FOR PAIN RELIEF AND PRESCRIPTION ANTIBIOTICS TO AVOID PROGRESSION TO OVERWHELMING SEPSIS.
ON (B)(6) 2022, THE CUSTOMER ALLEGED TO MEDELA LLC THAT THE DC JACK ON HER PUMP IN STYLE MAX FLOW IS BROKEN AND SHE IS STARTING TO GET MASTITIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 306428 | PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN) | PUMP, BREAST, POWERED | HGX | MEDELA LLC | 101041360 | 00020451401447 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Female | Other |