FDA Adverse Event Injury Summary report: N

VIVACIT-E®

MDR report key: 14605118 · Received June 6, 2022

Report

Report Number
14605118
Event Type
Injury
Date Received
June 6, 2022
Date of Event
May 6, 2022
Report Date
May 29, 2022
Manufacturer
ZIMMER, INC.
Product Code
LPH
UDI-DI
00889024572713
Adverse Event
Yes
Product Problem
Yes
Report Source
User Facility report
Reporter Location
AZ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Description of Event or Problem · 0

THE PATIENT HAD A BILATERAL TOTAL HIP ARTHROPLASTY (THA) DONE A FEW YEARS BACK AND HAD DONE WELL. PATIENT, SUBSEQUENTLY, HAD AN EXTENSIVE LUMBOSACRAL FUSION, AFTER WHICH POINT THE PATIENT DISLOCATING TOTAL HIP REPLACEMENT ANTERIORLY. CT SCANS AND ENGINEERS WITH THE CORIN GROUP OPTIMIZED POSITIONING SYSTEM (OPS) ANALYZED PATIENT'S HIPS IN THE SITTING AND STANDING POSITION AND GAVE RECOMMENDATIONS FOR THE IDEAL LOCATION OF THE ACETABULAR COMPONENT. SURGICAL/CLINICAL STAFF THEN WENT AHEAD AND DID THAT ACETABULAR REVISION APPROXIMATELY EIGHT WEEKS AGO. PATIENT HAD DONE WELL AND AT HER SIX-WEEK POSTOPERATIVE VISIT, PATIENT EXPRESSED ABSOLUTELY NO PAIN AND EVERYTHING WAS DOING VERY, VERY WELL. ABOUT ONE WEEK AGO, SOMETHING HAPPENED AND PATIENT DISLOCATED HER HIP. THERE WAS SOME CONCERN THAT THERE MAY HAVE BEEN MORE GOING ON THAN WHAT THE PATIENT SAID INITIALLY. HOWEVER, SHE SAID THAT SHE FLEXED HER HIP PAST 90 DEGREES IN THE SHOWER, ROTATED INTERNALLY AND THEN ADDUCTED HER FEMUR. THIS CAUSED THE FIRST DISLOCATION. OVER ONE WEEK SHE HAD TWO SUBSEQUENT DISLOCATIONS. BECAUSE OF THIS, WE DECIDED TO DO A REVISION. DESCRIPTION OF CATASTROPHIC FAILURE OF THE DUAL MOBILITY CONSTRUCT: AFTER THE FASCIA WAS OPENED, SURGICAL TEAM FOUND THE POLYETHYLENE LARGE BALL SITTING IN THE SOFT TISSUES INDEPENDENT OF THE OXINIUM FEMORAL HEAD, WHICH WAS STILL AFFIXED SUPERIORLY TO THE TRUNNION. THE REPRESENTATIVE IN THE ROOM WAS ASKED WHAT THE REPORTED INCIDENCE OF THIS WAS AND HE WAS AWARE THAT THERE HAD BEEN ONE REPORTED INCIDENCE OF THIS.

Description of Event or Problem · 0

THE PATIENT HAD A BILATERAL TOTAL HIP ARTHROPLASTY (THA) DONE A FEW YEARS BACK AND HAD DONE WELL. PATIENT, SUBSEQUENTLY, HAD AN EXTENSIVE LUMBOSACRAL FUSION, AFTER WHICH POINT THE PATIENT DISLOCATING TOTAL HIP REPLACEMENT ANTERIORLY. CT SCANS AND ENGINEERS WITH THE CORIN GROUP OPTIMIZED POSITIONING SYSTEM (OPS) ANALYZED PATIENT'S HIPS IN THE SITTING AND STANDING POSITION AND GAVE RECOMMENDATIONS FOR THE IDEAL LOCATION OF THE ACETABULAR COMPONENT. SURGICAL/CLINICAL STAFF THEN WENT AHEAD AND DID THAT ACETABULAR REVISION APPROXIMATELY EIGHT WEEKS AGO. PATIENT HAD DONE WELL AND AT HER SIX-WEEK POSTOPERATIVE VISIT, PATIENT EXPRESSED ABSOLUTELY NO PAIN AND EVERYTHING WAS DOING VERY, VERY WELL. ABOUT ONE WEEK AGO, SOMETHING HAPPENED AND PATIENT DISLOCATED HER HIP. THERE WAS SOME CONCERN THAT THERE MAY HAVE BEEN MORE GOING ON THAN WHAT THE PATIENT SAID INITIALLY. HOWEVER, SHE SAID THAT SHE FLEXED HER HIP PAST 90 DEGREES IN THE SHOWER, ROTATED INTERNALLY AND THEN ADDUCTED HER FEMUR. THIS CAUSED THE FIRST DISLOCATION. OVER ONE WEEK SHE HAD TWO SUBSEQUENT DISLOCATIONS. BECAUSE OF THIS, WE DECIDED TO DO A REVISION. DESCRIPTION OF CATASTROPHIC FAILURE OF THE DUAL MOBILITY CONSTRUCT: AFTER THE FASCIA WAS OPENED, SURGICAL TEAM FOUND THE POLYETHYLENE LARGE BALL SITTING IN THE SOFT TISSUES INDEPENDENT OF THE OXINIUM FEMORAL HEAD, WHICH WAS STILL AFFIXED SUPERIORLY TO THE TRUNNION. THE REPRESENTATIVE IN THE ROOM WAS ASKED WHAT THE REPORTED INCIDENCE OF THIS WAS AND HE WAS AWARE THAT THERE HAD BEEN ONE REPORTED INCIDENCE OF THIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306427 VIVACIT-E® PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS LPH ZIMMER, INC. 110031013 64689634 00889024572713

Patients

Seq Age Sex Outcome Treatment
1 22265 DA Female Hospitalization| O| R