FDA Adverse Event Malfunction Summary report: N

BD DIFCO¿ AGAR, GRANULATED

MDR report key: 14604647 · Received June 6, 2022

Report

Report Number
1119779-2022-00820
Event Type
Malfunction
Date Received
June 6, 2022
Date of Event
May 9, 2022
Report Date
July 1, 2022
Manufacturer
BECTON, DICKINSON & CO. (SPARKS)
Product Code
JSK
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THIS MDR SHOULD BE CONSIDERED CANCELLED. THIS PRODUCT WAS DETERMINED TO BE FOR LAB USE ONLY AND IS NOT SOLD AS A MEDICAL DEVICE IN THE US. THIS COMPLAINT HAS BEEN DETERMINED TO BE NOT REPORTABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD DIFCO¿ AGAR, GRANULATED THERE WAS ALTERED PHENOTYPIC GROWTH. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING ALTERED PHENOTYPIC GROWTH WITH THEIR 214530 AGAR FROM LOT 0139161. CUSTOMER STATES AGAR IS PRODUCING AN ALTERED GROWTH PHENOTYPE FOR ONE OF THEIR ORGANISMS. WHEN LB AND TSA PLATES MADE WITH THIS AGAR ARE USED TO CULTURE SALMONELLA ENTERICA SUBSP ENTERITIDIS, S. ENTERITIDIS DISPLAYS A STRESS-ASSOCIATED, FILAMENTOUS CELLULAR MORPHOLOGY (DETERMINED BY GRAM STAINING) AND IRREGULAR COLONY MORPHOLOGY (INCONSISTENT COLONY SIZE AND COLOR). CUSTOMER HAS TESTED A VARIETY OF POSSIBLE SOURCES, INCLUDING THE WATER, GLASSWARE, AUTOCLAVE, AND NUTRIENT SOURCE, AND ARE CONFIDENT IN OUR ASSESSMENT THAT THE ISSUE IS CAUSED BY THIS LOT OF DIFCO AGAR. USING A DIFFERENT LOT NUMBER OF THE SAME PRODUCT DOES NOT REPRODUCE THIS ISSUE. CUSTOMER ALSO TRIED OTHER ORGANISMS ON THE SAME MEDIA AND SAW TYPICAL GROWTH. CUSTOMER WANTED TO REACH OUT AND INFORM BD THAT THERE MAY BE AN ISSUE WITH THIS LOT. SINCE THE ISSUE WAS ONLY OBVIOUS WITH SOME SPECIES OF BACTERIA AND NOT OTHERS, THE ISSUE MIGHT BE HAVING ¿INVISIBLE¿ EFFECTS ON EXPERIMENTAL OUTCOMES, OR MAY HAVE RESULTED IN DATA THAT IS NOT RELIABLE OR REPLICABLE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING USE WITH BD DIFCO¿ AGAR, GRANULATED THERE WAS ALTERED PHENOTYPIC GROWTH. THERE WAS NO REPORT OF PATIENT IMPACT. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: CUSTOMER IS REPORTING ALTERED PHENOTYPIC GROWTH WITH THEIR 214530 AGAR FROM LOT 0139161. CUSTOMER STATES AGAR IS PRODUCING AN ALTERED GROWTH PHENOTYPE FOR ONE OF THEIR ORGANISMS. WHEN LB AND TSA PLATES MADE WITH THIS AGAR ARE USED TO CULTURE SALMONELLA ENTERICA SUBSP ENTERITIDIS, S. ENTERITIDIS DISPLAYS A STRESS-ASSOCIATED, FILAMENTOUS CELLULAR MORPHOLOGY (DETERMINED BY GRAM STAINING) AND IRREGULAR COLONY MORPHOLOGY (INCONSISTENT COLONY SIZE AND COLOR). CUSTOMER HAS TESTED A VARIETY OF POSSIBLE SOURCES, INCLUDING THE WATER, GLASSWARE, AUTOCLAVE, AND NUTRIENT SOURCE, AND ARE CONFIDENT IN OUR ASSESSMENT THAT THE ISSUE IS CAUSED BY THIS LOT OF DIFCO AGAR. USING A DIFFERENT LOT NUMBER OF THE SAME PRODUCT DOES NOT REPRODUCE THIS ISSUE. CUSTOMER ALSO TRIED OTHER ORGANISMS ON THE SAME MEDIA AND SAW TYPICAL GROWTH. CUSTOMER WANTED TO REACH OUT AND INFORM BD THAT THERE MAY BE AN ISSUE WITH THIS LOT. SINCE THE ISSUE WAS ONLY OBVIOUS WITH SOME SPECIES OF BACTERIA AND NOT OTHERS, THE ISSUE MIGHT BE HAVING ¿INVISIBLE¿ EFFECTS ON EXPERIMENTAL OUTCOMES, OR MAY HAVE RESULTED IN DATA THAT IS NOT RELIABLE OR REPLICABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
158647 BD DIFCO¿ AGAR, GRANULATED SUPPLEMENT, CULTURE MEDIA JSK BECTON, DICKINSON & CO. (SPARKS) 0139161

Patients

Seq Age Sex Outcome Treatment
1 Unknown