FDA Adverse Event Injury Summary report: N

IMP,TSV,4.1MM,DUAL SEL,HA

MDR report key: 14596508 · Received June 3, 2022

Report

Report Number
0002023141-2022-01387
Event Type
Injury
Date Received
June 3, 2022
Date of Event
December 26, 2021
Report Date
November 8, 2022
Manufacturer
ZIMMER DENTAL
Product Code
DZE
UDI-DI
00889024019300
PMA / PMN Number
K072589
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ZIMMER BIOMET COMPLAINT NUMBER (B)(4). WEIGHT UNKNOWN / NOT PROVIDED. A SUMMARY INVESTIGATION HAS BEEN COMPLETED FOR INFECTION EVENTS IDENTIFYING THAT A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED DUE TO THE WIDE RANGE OF EXTERNAL (NON-DESIGN/ NON-MANUFACTURING RELATED) FACTORS POTENTIALLY IMPACTING THE STERILITY OF THE IMPLANT AND ITS ENVIRONMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE INFECTION EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION AND DEVICE EVALUATION. DHR REVIEW WAS COMPLETED FOR THE SUBJECT LOT NUMBER 1233375. IT WAS CONFIRMED THAT ALL OPERATIONS AND INSPECTIONS WERE EXECUTED AS PER APPLICABLE PROCEDURE. NO DEVIATIONS OR NON-CONFORMANCES, WHICH COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT WAS NOTED AS PART OF THE DHR. LOT WAS INSPECTED AND PASSED ALL ACCEPTANCE CRITERIA BY QA. STERILIZATION RECORD OP# 11O - STR2 WAS REVIEWED AND VERIFIED TO HAVE PASSED ALL STERILIZATION ACTIVITIES WITH NO ISSUES OR NONCONFORMITIES IDENTIFIED. COMPLAINT HISTORY REVIEW WAS PERFORMED FOR THE REPORTED LOT NUMBER 1233375 FOR SIMILAR EVENTS AND NO OTHER COMPLAINT WAS IDENTIFIED. AS DOCUMENTED IN THE SUMMARY INVESTIGATION, CONTRIBUTING FACTORS FOR THESE REPORTED EVENTS LIKELY EXIST OUTSIDE OF ZIMMER BIOMET CONTROL, INCLUDING THOSE RELATED TO PATIENT BIOLOGICAL FACTORS/CONDITION AND SURGICAL TECHNIQUE. BASED ON THE SUMMARY INVESTIGATION, NO MALFUNCTION OCCURRED UPON INVESTIGATION. THE REPORTED EVENT REMAINS NON-VERIFIABLE AS IT IS A MEDICAL CONDITION.

Description of Event or Problem · 0

DOCTOR INDICATED INFECTION AT TOOTH SITE #45. THE PATIENT MANIFESTS PAIN IN THE AREA OF THE IMPLANT AND A CLINICAL EVALUATION IS CARRIED OUT. IT PRESENTS PAIN TO THE TOUCH AND NO FIXATION OF THE IMPLANT IS OBSERVED AND IT IS PROCEEDED TO EXTRACT.

Description of Event or Problem · 0

NO ADDITIONAL EVENT INFORMATION RECEIVED AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1587747 IMP,TSV,4.1MM,DUAL SEL,HA DENTAL IMPLANT DZE ZIMMER DENTAL TSV4H10 1233375 00889024019300

Patients

Seq Age Sex Outcome Treatment
1 48 YR Male Required Intervention