FDA Adverse Event Death Summary report: N

ENROUTE TRANSCAROTID STENT SYSTEM

MDR report key: 14593450 · Received June 3, 2022

Report

Report Number
3014526664-2022-00076
Event Type
Death
Date Received
June 3, 2022
Date of Event
April 17, 2022
Report Date
June 3, 2022
Manufacturer
SILK ROAD MEDICAL INC.
Product Code
NIM
UDI-DI
00811311020508
PMA / PMN Number
P140026
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SECONDARY MDR WAS REPORTED UNDER 3014526664-2022-00077 AS THERE ARE TWO STENTS ASSOCIATED WITH THE SAME EVENT/PATIENT. THE PRODUCT ASSOCIATED WITH THIS COMPLAINT WAS NOT RETURNED TO THE MANUFACTURER FOR ANALYSIS. A REVIEW OF THE MANUFACTURING RECORDS FOR THIS DEVICE WAS COMPLETED AND NO ISSUES WERE IDENTIFIED THAT COULD HAVE LED TO THE ADVERSE EVENT REPORTED. THERE IS NO INDICATION THAT A MALFUNCTION OF THE SRM DEVICE OCCURRED; THE CAUSE OF THE POST-OPERATIVE COMPLICATION IS UNKNOWN, THEREFORE, THE COMPLAINT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION. COMPLAINTS WILL CONTINUE TO BE REVIEWED AND MONITORED FOR TRENDS. A SUPPLEMENTAL MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 0

IT WAS REPORTED THAT TWO DAYS AFTER COMPLETION OF A LEFT TRANSCAROTID ARTERY REVASCULARIZATION (TCAR) PROCEDURE, THE PATIENT'S STENT THROMBOSED. A STROKE OCCURRED, AND THE PATIENT WAS PARALYZED ON THE RIGHT LEG AND ARM. THE PHYSICIAN PERFORMED A THROMBECTOMY AND BALLOON ANGIOPLASTY AND PLACED THE PATIENT ON TICAGRELOR/BRILINTA. ADDITIONAL INFORMATION RECEIVED ON (B)(6) 2022 INDICATED THAT PATIENT WAS IN THE HOSPITAL FOR 30 DAYS AND THEN WAS DISCHARGED. THE PATIENT WENT TO A REHABILITATION FACILITY AND SUBSEQUENTLY PASSED AWAY. THE CAUSE AND DATE OF DEATH ARE UNKNOWN. AT THIS TIME, THERE IS NO INDICATIONS AS TO WHAT MAY HAVE CAUSED THE THROMBUS, OR IF THE REPORTED FAILURE IS RELATED TO PROCEDURAL ISSUES, PATIENT RESISTANCE TO DUAL ANTIPLATELET THERAPY (DAPT) MEDICATION, OR A SILK ROAD MEDICAL DEVICE FAILURE, HENCE, THE EVENT WILL BE REPORTED OUT OF ABUNDANCE OF CAUTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1586408 ENROUTE TRANSCAROTID STENT SYSTEM ENROUTE TSS NIM SILK ROAD MEDICAL INC. SR-094-CS 18034547 00811311020508

Patients

Seq Age Sex Outcome Treatment
1 77 YR Male Death| S| R