FDA Adverse Event Malfunction Summary report: N

TRIAGE PROFILER SOB 25 TEST KIT

MDR report key: 1459070 · Received June 25, 2009

Report

Report Number
2027969-2009-00496
Event Type
Malfunction
Date Received
June 25, 2009
Date of Event
June 2, 2009
Report Date
June 25, 2009
Manufacturer
BIOSITE INCORPORATED
Product Code
NBC
PMA / PMN Number
042723
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RETAINS FROM DEVICE LOT W44978 WERE TESTED USING IN-HOUSE CALIBRATOR (CAL H). RESULTS ARE AS FOLLOW: NO ERROR CODES OR STANDARD OUTLIERS WERE OBSERVED. PRODUCT SUPPORT OBSERVED ALL DATA POINTS TO BE WITHIN THE MANUFACTURING 1SD RANGE. NO LOW OR HIGH VALUES WERE OBSERVED. COMPLAINT NOT CONFIRMED. NO DISCREPANT RESULTS WERE OBSERVED. THE CUSTOMER DID NOT RETURN ANY SAMPLE FOR THIS COMPLAINT. PRODUCT SUPPORT (PS) WAS UNABLE TO VERIFY THE CUSTOMER'S COMPLAINT. PS COULD NOT RULE OUT ANY SAMPLE SPECIFIC OR ENVIRONMENTAL FACTOR THAT MAY AFFECT ANALYTE RECOVERY. PS COULD NOT CONFIRM THE CLIENT'S OBSERVATION WITHOUT RETURN OF PATIENT SAMPLE. RETAIN TESTING OF THE DEVICE LOT AGAINST INTERNAL PLASMA CONTROLS SHOWED BNP RESULTS WITH NO 2SD OUTLIERS. RECOVERY WAS 95% WITH A %CV OF 4.7. A REVIEW OF MANUFACTURING (MFG) RELEASED DATA FOR THE DEVICE LOT SHOWED RESULTS WITHIN MFG SPECIFICATIONS. PRODUCT DEFICIENCY WAS NOT ESTABLISHED. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER REPORTING CUSTOMER WITH DISCREPANT BNP RESULTS ON TRIAGE PROFILER SOB 25 TEST KIT. BNP RUN TODAY GAVE 202 ON TRIAGE SOB. SAMPLE SENT TO LAB, UNKNOWN METHOD RUN WITH BNP RESULT OF 4100. RERUN OF SAME SAMPLE IN ED ON DIFFERENT TRIAGE METER RESULTED IN BNP OF 2100, ON ORIGINAL METER BNP WAS 4000. NO INFORMATION ON PATIENT PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TRIAGE PROFILER SOB 25 TEST KIT CARDIAC MARKER TEST NBC BIOSITE INCORPORATED 97300 W44978

Patients

Seq Age Sex Outcome Treatment
1 NI