MTS ANTI-IGG, -C3D CARD
Report
- Report Number
- 1056600-2009-00097
- Event Type
- Malfunction
- Date Received
- May 8, 2009
- Date of Event
- April 14, 2009
- Report Date
- May 7, 2009
- Manufacturer
- MICRO TYPING SYSTEMS, INC,
- Product Code
- CFQ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
SATISFACTORY RESULTS WERE OBTAINED AT OCD USING RETAINED CARDS FROM MTS ANTI-IGG - C3D CARD LOT # 071608005-04 AND MTS ANTI-IGG CARD LOT # 112408001-30 USING KNOWN DAT POSITIVE AND NEGATIVE SAMPLES. THE GEL CARDS PERFORMED AS EXPECTED AT OCD. THE CUSTOMER'S COMPLAINT WAS NOT CONFIRMED. THE SAMPLE IN QUESTION WAS SUBMITTED TO OCD FOR ADD'L TESTING. A DEFINITE ROOT CAUSE WAS NOT DETERMINED REGARDING THE NEGATIVE RESULTS OBTAINED AT THE CUSTOMER SITE. A COMPLAINT REVIEW INDICATED NO OTHER COMPLAINTS HAVE BEEN LOGGED AGAINST LOT# 071608005-04. BATCH RECORD REVIEW WAS WITHIN RELEASE SPECIFICATIONS. INCIDENT IS ISOLATED. (B) (4).
THE CUSTOMER REPORTED THAT A DAT POSITIVE SAMPLE FAILED TO REACT IN THE MTS ANTI-IGG, -C3D CARD LOT# 071608005-04. THE SAMPLE REACTED POSITIVE (1+) USING THE MTS ANTI-IGG CARD. A FALSE NEGATIVE DAT MAY LEAD TO CONFUSION IN THE DIAGNOSIS OF ANEMIA OR MISDIAGNOSIS OF A HEMOLYTIC TRANSFUSION REACTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MTS ANTI-IGG, -C3D CARD | ANTI-HUMAN GLOBULIN REAGENT | CFQ | MICRO TYPING SYSTEMS, INC, | NA | 071608005-04 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |