FDA Adverse Event Malfunction Summary report: N

MTS ANTI-IGG, -C3D CARD

MDR report key: 1458586 · Received May 8, 2009

Report

Report Number
1056600-2009-00097
Event Type
Malfunction
Date Received
May 8, 2009
Date of Event
April 14, 2009
Report Date
May 7, 2009
Manufacturer
MICRO TYPING SYSTEMS, INC,
Product Code
CFQ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SATISFACTORY RESULTS WERE OBTAINED AT OCD USING RETAINED CARDS FROM MTS ANTI-IGG - C3D CARD LOT # 071608005-04 AND MTS ANTI-IGG CARD LOT # 112408001-30 USING KNOWN DAT POSITIVE AND NEGATIVE SAMPLES. THE GEL CARDS PERFORMED AS EXPECTED AT OCD. THE CUSTOMER'S COMPLAINT WAS NOT CONFIRMED. THE SAMPLE IN QUESTION WAS SUBMITTED TO OCD FOR ADD'L TESTING. A DEFINITE ROOT CAUSE WAS NOT DETERMINED REGARDING THE NEGATIVE RESULTS OBTAINED AT THE CUSTOMER SITE. A COMPLAINT REVIEW INDICATED NO OTHER COMPLAINTS HAVE BEEN LOGGED AGAINST LOT# 071608005-04. BATCH RECORD REVIEW WAS WITHIN RELEASE SPECIFICATIONS. INCIDENT IS ISOLATED. (B) (4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT A DAT POSITIVE SAMPLE FAILED TO REACT IN THE MTS ANTI-IGG, -C3D CARD LOT# 071608005-04. THE SAMPLE REACTED POSITIVE (1+) USING THE MTS ANTI-IGG CARD. A FALSE NEGATIVE DAT MAY LEAD TO CONFUSION IN THE DIAGNOSIS OF ANEMIA OR MISDIAGNOSIS OF A HEMOLYTIC TRANSFUSION REACTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MTS ANTI-IGG, -C3D CARD ANTI-HUMAN GLOBULIN REAGENT CFQ MICRO TYPING SYSTEMS, INC, NA 071608005-04

Patients

Seq Age Sex Outcome Treatment
1 Other