HYGENIC
Report
- Report Number
- 2416455-2022-00008
- Event Type
- Injury
- Date Received
- June 2, 2022
- Date of Event
- May 3, 2022
- Report Date
- June 1, 2022
- Manufacturer
- COLTENE WHALEDENT INC.
- Product Code
- EEF
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PL
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
INVESTIGATION: THIS COMPLAINT IS FOR MATERIAL H09960 #7 WINGED FIESTA CLAMP, BATCH K96855. INVENTORY ANALYSIS SHOWS NO REMAINING PRODUCT IN STOCK OF THIS BATCH NUMBER. SAP SHOWS THE STOCK OF THIS BATCH OF 2,000 WAS DEPLETED ON 2/24/2022. THE CLAMP WAS RETURNED FOR INVESTIGATION. THE CLAMP HAS A DATE CODE OF ND WHICH MEANS IT WAS MANUFACTURED IN 2018 JUL/AUG. THERE ARE MULTIPLE LARGE RUST SPOTS ON THE CLAMP THAT CAN BE SEEN WITH THE NAKED EYE. THE CLAMP WAS ALSO EXAMINED UNDER A MICROSCOPE. THE SURFACE OF THE CLAMP IS COVERED IN A WHITE RESIDUE. THE CLAMP BROKE ACROSS ONE OF THE RUST SPOTS. THERE ARE ALSO SMALL RUST SPOTS ON THE BREAK SURFACE. THE CUSTOMER CLEANS THEIR CLAMPS USING PRIMADENT PROTECT AND STERILIZATION FOR 30 MINUTES IN 132 °C. THE PH VALUE OF THIS CLEANING SOLUTION IS >11. THIS CLEANING PROCEDURE BEING USED BY THE CUSTOMER IS NOT THE CLEANING PROCEDURE STATED IN THE COLTENE IFU. THE IFU SPECIFICALLY STATES A PH NEUTRAL CLEANING SOLUTION SHOULD BE USED. ALSO, THE CLAMPS SHOULD BE STERILIZED FOR 15 MINUTES AT 132°C USING A GRAVITY AUTOCLAVE OR 4 MINUTES AT 132°C USING A PRE-VACUUM AUTOCLAVE. THE PRIMADENT PROTECT SOLUTION AND THE HIGH STERILIZATION TIME MAY HAVE A NEGATIVE IMPACT ON THE PRODUCT LIFE AND LIKELY ATTRIBUTED TO THE EARLY DEGRADATION AND BREAKAGE OF THE CLAMP. ROOT CAUSE: BREAK CAUSED BY MATERIAL DEGRADATION. MATERIAL DEGRADATION DUE TO CUSTOMER CLEANING PROCEDURE.
ON 5/3/2022 THE CUSTOMER STATED THE FOLLOWING: DENTAL CLAMPS BROKEN AFTER 2 USES IN PATIENT MOUTH. WASHING IN SOLUTION. DISINFECTED IN PRIMADENT PROTECT. STERILIZATION 30 MINUTES IN 132. USED ONLY ONE 2 TIME ( SINCE PURCHASED 10/2021). NOBODY WAS INJURED. NO SURGICAL INTERVENTION. BECAUSE A SIMILAR EVENT WAS PREVIOUSLY REPORTED WHERE THE DENTAL CLAMP BROKE DURING TREATMENT AND THE DENTAL DAM CAME OFF OF THE TOOTH CAUSING SALIVA TO INTRUDE AND TREATMENT TO BE RESTARTED, IF THIS MALFUNCTION WERE TO RECUR, AN INJURY COULD HAPPEN. OUT OF AN ABUNDANCE OF CAUTION AND PER 21 CFR 803, THIS EVENT IS BEING REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2161119 | HYGENIC | DENTAL DAM CLAMP | EEF | COLTENE WHALEDENT INC. | K96855 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown | Required Intervention |