FDA Adverse Event Malfunction Summary report: N

ORTHO SUMMIT SAMPLE HANDLING SYSTEM

MDR report key: 1456894 · Received August 14, 2009

Report

Report Number
2250051-2009-00255
Event Type
Malfunction
Date Received
August 14, 2009
Date of Event
August 11, 2009
Report Date
August 14, 2009
Manufacturer
HAMILTON BONADUZ AG
Product Code
JTC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT HAS BEEN INVESTIGATED. THE FIELD SERVICE ENGINEER (FSE) VERIFIED ALL TIP HOLD AND PULL FORCES MET SPECIFICATION. FSE FOUND X POSITION TIP TAKE UP OFF BY ONE STEP AND ALIGNED TIP TAKE UP. THE INSTRUMENT WAS TESTED WITHOUT FURTHER PROBLEM. INSTRUMENT IS OPERATING AS EXPECTED AND NO REPAIRS WERE NECESSARY.

Description of Event or Problem · 1

THE ORTHO SUMMIT SAMPLE HANDLING SYSTEM INSTRUMENT DID NOT PIPETTE SAMPLE AND REAGENT AND DID NOT GENERATE AN ERROR MESSAGE. NO DEATH OR SERIOUS INJURY WAS ASSOCIATED WITH THIS INCIDENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC. (B) (4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ORTHO SUMMIT SAMPLE HANDLING SYSTEM SAMPLE PIPETTER JTC HAMILTON BONADUZ AG 7003009

Patients

Seq Age Sex Outcome Treatment
1