FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 14567260 · Received June 1, 2022

Report

Report Number
3008642652-2022-20092
Event Type
Malfunction
Date Received
June 1, 2022
Date of Event
May 18, 2022
Report Date
June 1, 2022
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 107) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO A DEFECTIVE U104 PXA PROCESSOR ON THE COMPUTER/ANALOG BOARD. THE ROOT CAUSE FOR THE DEFECTIVE U104 PXA PROCESSOR COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR.

Description of Event or Problem · 0

A US DISTRIBUTOR REPORTED THAT A MONITOR WILL NOT POWER UP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308091 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 Unknown