FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 14567260
·
Received June 1, 2022
Report
- Report Number
- 3008642652-2022-20092
- Event Type
- Malfunction
- Date Received
- June 1, 2022
- Date of Event
- May 18, 2022
- Report Date
- June 1, 2022
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 0
DEVICE EVALUATION OF MONITOR HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 107) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO A DEFECTIVE U104 PXA PROCESSOR ON THE COMPUTER/ANALOG BOARD. THE ROOT CAUSE FOR THE DEFECTIVE U104 PXA PROCESSOR COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DAMAGED MONITOR.
Description of Event or Problem · 0
A US DISTRIBUTOR REPORTED THAT A MONITOR WILL NOT POWER UP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 308091 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL MANUFACTURING CORPORATION | LIFEVEST WCD 4000 SYSTEM |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |