FDA Adverse Event
Malfunction
Summary report: N
UNI-VENT, MODEL 731
MDR report key: 14567095
·
Received June 1, 2022
Report
- Report Number
- 1220908-2022-01953
- Event Type
- Malfunction
- Date Received
- June 1, 2022
- Report Date
- May 11, 2022
- Manufacturer
- ZOLL MEDICAL CORPORATION
- Product Code
- CBK
- PMA / PMN Number
- K111473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
Additional Manufacturer Narrative · 0
ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.
Additional Manufacturer Narrative · 0
THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO AN INTEGRATED CIRCUIT ON THE CENTRAL PROCESSING UNIT BOARD (CPU). THE CENTRAL PROCESSING UNIT BOARD WAS REPLACED TO RESOLVE THE CUSTOMER'S REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.
Description of Event or Problem · 0
COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1426605 | UNI-VENT, MODEL 731 | VENTILATOR | CBK | ZOLL MEDICAL CORPORATION | 799-EMVP-01-01 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Unknown |