FDA Adverse Event Malfunction Summary report: N

UNI-VENT, MODEL 731

MDR report key: 14567095 · Received June 1, 2022

Report

Report Number
1220908-2022-01953
Event Type
Malfunction
Date Received
June 1, 2022
Report Date
May 11, 2022
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
CBK
PMA / PMN Number
K111473
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO ZOLL MEDICAL CORPORATION; THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO AN INTEGRATED CIRCUIT ON THE CENTRAL PROCESSING UNIT BOARD (CPU). THE CENTRAL PROCESSING UNIT BOARD WAS REPLACED TO RESOLVE THE CUSTOMER'S REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS FOR REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING A ROUTINE SHIFT CHECK BY A CLINICIAN, THE DEVICE POWERED ON WITHOUT DISPLAY. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1426605 UNI-VENT, MODEL 731 VENTILATOR CBK ZOLL MEDICAL CORPORATION 799-EMVP-01-01 NA

Patients

Seq Age Sex Outcome Treatment
1 Unknown