HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP
Report
- Report Number
- 3007042319-2022-05285
- Event Type
- Injury
- Date Received
- June 1, 2022
- Date of Event
- April 28, 2022
- Report Date
- August 3, 2022
- Manufacturer
- HEARTWARE, INC.
- Product Code
- DSQ
- UDI-DI
- 00888707003261
- PMA / PMN Number
- P100047
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION OF THIS EVENT IS PENDING AND A SUPPLEMENTAL REPORT WILL BE SENT UPON ITS COMPLETION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
A SUPPLEMENTAL REPORT IS BEING SUBMITTED FOR INVESTIGATION COMPLETION. PRODUCT EVENT SUMMARY: THE VENTRICULAR ASSIST DEVICE (VAD) WAS NOT RETURNED FOR EVALUATION. THE REPORTED LOW FLOW EVENT COULD NOT BE CONFIRMED VIA REVIEW OF THE CONTROLLER LOG FILES SINCE LOG FILES COVERING THE REPORTED EVENT DATE WERE NOT AVAILABLE FOR ANALYSIS. INFORMATION RECEIVED FROM THE SITE INDICATED THAT IN ADDITION TO THE LOW FLOWS THE PATIENT EXPERIENCED ANEMIA AND LOW HEMOGLOBIN, AND AN UPPER GASTROINTESTINAL (GI) BLEED WAS SUSPECTED. A COLONOSCOPY WAS PERFORMED WITH NO FINDINGS. AN ESOPHAGOGASTRODUODENOSCOPY (EGD) WAS PERFORMED, AND THE PATIENT WAS TREATED WITH ARGON PLASMA COAGULATION (APC). BASED ON THE AVAILABLE INFORMATION, THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE REPORTED EVENT. BASED ON THE RISK DOCUMENTATION, POSSIBLE CAUSES OF THE REPORTED LOW FLOW EVENT MAY BE ATTRIBUTED TO MULTIPLE FACTORS INCLUDING BUT NOT LIMITED TO THROMBUS AT THE INFLOW CANNULA/OUTFLOW GRAFT, CONSTRICTION AT THE OUTFLOW GRAFT, POOR VAD FILLING, AND/OR INAPPROPRIATE PUMP ROTATIONAL SPEED. PER THE INSTRUCTIONS FOR USE, GI BLEEDING IS A KNOWN POTENTIAL COMPLICATION ASSOCIATED WITH THE IMPLANTATION OF A VAD. THERE WAS NO EVIDENCE THAT THE PATIENT HAD A HISTORY OF BLEEDING EVENTS. POSSIBLE CLINICAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT INCLUDE THE PATIENT¿S PRE-EXISTING HISTORY AND RELATED COMORBIDITIES, THE PROGRESSION OF THEIR UNDERLYING DISEASE, ISSUES RELATED TO THE THERAPEUTIC USE OF ANTICOAGULANT AND ANTIPLATELET MEDICATIONS AND THE PATIENT'S COMPLEX POST-OPERATIVE COURSE. THERE ARE POSSIBLE PATIENT, PHARMACOLOGICAL AND PROCEDURAL FACTORS THAT MAY HAVE CONTRIBUTED TO THIS EVENT. INVESTIGATION OF THIS EVENT IS COMPLETED AND THE FILE WILL BE CLOSED. IF NEW INFORMATION IS RECEIVED, THE FILE WILL BE RE-OPENED AND A SUPPLEMENTAL WILL BE SUBMITTED. MEDTRONIC IS SUBMITTING THIS REPORT TO COMPLY WITH FDA REPORTING REGULATIONS UNDER 21 CFR PARTS 4 AND 803. THIS REPORT IS BASED UPON INFORMATION OBTAINED BY MEDTRONIC, WHICH THE COMPANY MAY NOT HAVE BEEN ABLE TO FULLY INVESTIGATE OR VERIFY PRIOR TO THE DATE THE REPORT WAS REQUIRED BY THE FDA. MEDTRONIC HAS MADE REASONABLE EFFORTS TO OBTAIN MORE COMPLETE INFORMATION AND HAS PROVIDED AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. IN PARTICULAR, THIS REPORT DOES NOT CONSTITUTE AN ADMISSION BY ANYONE THAT THE PRODUCT DESCRIBED IN THIS REPORT HAS ANY ¿DEFECTS¿ OR HAS ¿MALFUNCTIONED¿. THESE WORDS ARE INCLUDED IN THE FDA 3500A FORM AND ARE FIXED ITEMS FOR SELECTION CREATED BY THE FDA TO CATEGORIZE THE TYPE OF EVENT SOLELY FOR THE PURPOSE OF REGULATORY REPORTING. MEDTRONIC OBJECTS TO THE USE OF THESE WORDS AND OTHERS LIKE THEM BECAUSE OF THE LACK OF DEFINITION AND THE CONNOTATIONS IMPLIED BY THESE TERMS. THIS STATEMENT SHOULD BE INCLUDED WITH ANY INFORMATION OR REPORT DISCLOSED TO THE PUBLIC UNDER THE FREEDOM OF INFORMATION ACT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. A GOOD FAITH EFFORT WILL BE MADE TO OBTAIN THE APPLICABLE INFORMATION RELEVANT TO THE REPORT. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
IT WAS REPORTED THAT THE VENTRICULAR ASSIST DEVICE (VAD) PATIENT WAS ADMITTED FOR CONTINUOUS LOW FLOWS DUE TO ANEMIA AND LOW HEMOGLOBIN (HGC) AND AN UPPER GASTROINTESTINAL BLEED (GIB) WAS SUSPECTED. THEIR HGC WAS 7.7 WHICH WAS DOWN FROM 15.1 TWO MONTHS PRIOR. THEIR INTERNATIONAL NORMALIZED RATIO (INR) WAS 3.0 WHICH WAS DOWN FROM 4.1 A FEW DAYS PRIOR WITH A GOAL OF 2.5-3.5. A COLONOSCOPY FOUND NOTHING AND ESOPHAGOGASTRODUODENOSCOPY (EGD) SPOT TREATED WITH ARGON PLASMA COAGULATION. THE PATIENT WAS NOT TRANSFUSED AS THEIR HGC TRENDED BACK UP AND THEY WERE RELISTED FOR TRANSPLANT WITH THEIR BODY MASS INDEX LESS THAN 40 AT STATUS 4. THE VAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1462069 | HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP | VENTRICULAR (ASSISST) BYPASS | DSQ | HEARTWARE, INC. | 1103 | 00888707003261 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR | Male | Required Intervention| H |