FDA Adverse Event Injury Summary report: N

ROHO® QUADTRO SELECT® HIGH PROFILE® CUSHION

MDR report key: 14565550 · Received June 1, 2022

Report

Report Number
3008630266-2022-00003
Event Type
Injury
Date Received
June 1, 2022
Date of Event
March 1, 2022
Report Date
June 1, 2022
Manufacturer
ROHO, INC.
Product Code
KIC
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

THE CUSTOMER STATED HE HAS A PRESSURE INJURY DUE TO CUSHION GOING FLAT IN REAR LEFT QUAD AREA, BUT ROHO, INC. HAS NOT SEEN MEDICAL RECORDS TO CONFIRM THIS. THE CUSHION WAS RETURNED AND THE EVALUATION REPORT IS ATTACHED. NO PROBLEM FOUND DURING EVALUATION THE CUSHION. ISSUE DETERMINED TO BE INCOMPATIBILITY WITH USER'S ANATOMY AND ISOFLO VALVE ON CUSHION. IF ROHO, INC. BECOMES AWARE OF ADDITIONAL, RELEVANT INFORMATION, A FOLLOW UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

CUSTOMER CLAIMS THAT LEAK IN REAR LEFT QUAD OF CUSHION RESULTED IN A PRESSURE INJURY. USER DEVELOPED A PRESSURE INJURY IN (B)(6) 2022.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1933850 ROHO® QUADTRO SELECT® HIGH PROFILE® CUSHION WHEELCHAIR CUSHION KIC ROHO, INC. QS99C

Patients

Seq Age Sex Outcome Treatment
1 48 YR Male Other